2.3 Control unit functions – touchscreen

2.3.1. Automatic setting (1)

The mattress system regulates the mattress’s 

internal pressure independently and without 

manual adjustment to different values based on the user’s 

weight and position. No manual action is required to affect 

the internal pressure of the mattress. This function works in 

the following three ways: 
1. The system performs an automatic setting when the 

system has been started.
2. The system itself will start this function if any significant 

changes occur during use.
3. The system performs an automatic setting at fixed 

intervals even if no significant changes have occurred.
Once automatic setting is complete, the system will revert to

the previously used program. At first start‐up, the setting is

the basic setting of Pulsating mode, 10 minutes.

2.3.2. Comfort settings (2)

The pressure can be increased in two steps de‐

pending on the user’s comfort requirements. This 

increase is done based on the automatic 

setting performed by the control unit. (The control 

unit automatically sets the optimised pressure for the user.)

2.3.3. Programs (3)

Three programs are available. Press the button to switch 

between programs.

1. Alternating mode.

 Here, the pressure 

alternates between the cells. The cycle period 

can be changed according to the user’s wishes 

and requirements. Choose between 10, 15, 20 or 

25 minutes. The longer the cycle period, the  

slower the alternations. We recommend a basic setting of 10 

minutes.

2. Pulsating mode.

 A comfort mode that 

combines Constant low pressure with 

Alternating mode. The cycle period can be 

changed according to the user’s wishes and  

 requirements. The longer the cycle period, the  slower the 

alternations. We recommend a basic setting of 10 minutes.

3. Constant low pressure.

 Here, the pressure 

will be same in all cells. This is also called ‘low 

pressure’.

2.3.4. Maximum pressure (Caring mode) (4)

With this function, the entire mattress is 

inflated and provides fixed support. This 

function automatically switches to the preced‐

ing setting after approx. 20 minutes. This function should be 

used when nursing the user, shifting the user’s position or 

moving the user in or out of bed.

2.3.5. Screen lock (5)

Press the Screen Lock button to lock or unlock the 

screen. The display indicates when the screen has been 

locked. The screen locks automatically if left untouched for 

five minutes. This is to prevent the settings being changed 

accidentally.

To unlock, press the padlock          and confirm using

again.

2.3.6. Information (6)

Button that takes you to the information menu.

In the information menu you will find:

.

 Information about the product

.

 Contact information for Care of Sweden

.

 Shortcut to quick tour

.

 Deflation function

2.3.7. Battery indicator (7)

If the control unit has been fitted with a backup battery 

(variant), the condition of the battery is shown as follows:

2.3.8. Notifications (8)

Notifications are shown by           in the upper left corner 

after acknowledgement (deactivation). Find out more about 

notifications in section 4.6.

(8)

(2)

(3)

(5)

(1)

(7)

(4)

(6)

NOTE:

The automatic settings are disengaged when using this pro‐

gram.

Battery 100%

Battery 75%

Battery 50%
Battery <25%. Charge immediately 

(control unit 

prepares constant low pressure therapeutic mode).
Battery charging

NOTE:

To ensure optimum function and condition, the backup 

battery should never be discharged completely. The battery 

charges up when the control unit is operating. A fully charged 

battery lasts for at least five hours.
If the control unit is left without operating for a long period, 

the backup battery should be charged every six months for 

maintenance purposes.

NOTE:

During battery operation, it is possible to select Constant low 

pressure and Pulsating mode, but it is not possible to choose 

the cycle period. The cycle period will be 20 minutes (or longer 

if this was set before the transition to battery operation).

7

Содержание CuroCell UNO

Страница 1: ...Instructions for use 2021 06 09 CuroCell UNO Automatic air mattress systems Instructions for use item number 95 001401 EN0000 ...

Страница 2: ... 2 Product description 5 3 Assembly installation 8 4 Operation 8 5 Reuse cleaning and reconditioning 12 6 Storage 12 7 Maintenance 13 8 Troubleshooting 13 9 Technical description 14 10 Spare parts 17 11 Options and accessories 18 12 Warranty 18 13 Other information 19 2 ...

Страница 3: ...the following 11 Do not use the product in bathrooms or other area where there is a risk of the control unit coming into contact with water or other liquids Except for specified cleaning never handle a product that has come into contact with water liquid Pull the plug out of the electrical socket immediately and send the product to an authorized service technician for servicing 12 Do not use close...

Страница 4: ... diagnosis and general condition The mattress should only be used lying in the lengthwise direction on the mat tress with the feet at the end marked with the feet symbol No training or special skills are required for the use of Curo Cell A4 It is designed to be easy to handle and understand reading the quick guide and the instructions for use before use is recommended Prescription shall be made by...

Страница 5: ... and mattress 1 Mattress 2 CPR Quick deflation 3 Tube connection control unit 4 Control unit 1 2 3 4 Mattress Hygienic cover top part Cell holder integrated in inner cover Cells Safety mattress and tubing set Inner cover Cover bottom part 5 ...

Страница 6: ... cover It is easy to handle and maintain and conforms to the stringent requirements in respect of cleaning and hygiene The hygiene cover is vapour permeable i e the vapour is transported away thus reducing the risk of skin maceration Stone Cover A removable cover with a drip proof zip fastener on four sides and welded seams 2 2 Control unit 1 Tube CPR connection 4 Power switch On Off 1 0 8 Power a...

Страница 7: ...led low pressure 2 3 4 Maximum pressure Caring mode 4 With this function the entire mattress is inflated and provides fixed support This function automatically switches to the preced ing setting after approx 20 minutes This function should be used when nursing the user shifting the user s position or moving the user in or out of bed 2 3 5 Screen lock 5 Press the Screen Lock button to lock or unloc...

Страница 8: ... is found contact Care of Sweden or your local distributor before using the product NOTE Equipment with an electrical safety class of II must always be used for home care and in nursing homes see section 9 3 Labelling In hospitals and equivalent sites equipment with an electrical safety class of I or II may be used depending on whether or not the socket is earthed See section 9 1 Technical specifi...

Страница 9: ...re that the mattress is working properly This must be performed regularly we recommend once per shift or every eight hours Open the cover and insert a hand between the upper cells beneath the user s pelvic region centre of the mattress Check to ensure that there is a gap to the underlying mattress so that the user does not bottom out We recommend a gap of one or two finger widths If the gap is too...

Страница 10: ... control unit Notifications Notifications are shown with a warning triangle on the main menu with a dedicated display image or simply with a sound and flashing orange light in the function button on the front Notifications are silenced and acknowledged disabled using the function button on the front In the event of problems with Notifications see section 8 for troubleshooting Silent notifications ...

Страница 11: ...power adaptor can be removed from the wall socket and the control unit left over the end of the bed during transport For optimum comfort Care of Sweden recommends first placing the control unit in Constant low pressure mode and waiting for 5 minutes before disconnecting The mattress will retain air for at least eight hours If the control unit has a backup battery the control unit will automaticall...

Страница 12: ...ch as alcohol with or without tenside surfactants Isopropanol 70 oxidizing solutions chlorine max 1 or hydrogen peroxide max 1 5 5 2 Reconditioning CONTROL UNIT See section 5 1 for instructions MATTRESS 1 Disconnect the tube connector from the control unit and remove the air from the mattress 2 Remove the covers For more information of cleaning the covers see section 5 1 3 Move the mattress to a c...

Страница 13: ... user is bottoming out The mattress moves around Some cells have less air The control unit makes a noise vibrations can be felt Solution Check that the power adaptor has been connected to the mains supply Check that the LED on the power adaptor is showing green Check that the power adaptor has not become loose both the main supply and the control unit Check that the switch is set to 1 On If the co...

Страница 14: ...0 C Storage 20 C 45 C Transport 25 C 70 C Operation 15 93 non condensing Storage 93 non condensing 700 hPa 1060 hPa 10 x Ni MH 1 2 V 3500 mAh Class I Class II Type BF IP42 The device is not intended for use with flammable anaesthetic gases Mattress SPECIFICATION CuroCell UNO 80 85 90 105 120 x 200 210 220 x 20 cm 11 kg 80x200 cm Pulsating and Alternating Ungrounded AC outlet electrical safety clas...

Страница 15: ... key THE MATTRESS The mattress cover and inner cover are marked as shown below example For an explanation see section 9 4 Symbol key WARNING To avoid the risk of electric shock equipment marked as being in electrical safety class I must be connected to an earthed socket only Double insulation symbol denotes electrical safety class II may be connected to an unearthed socket Earth symbol denotes ele...

Страница 16: ...rectly on the bed base Counteracts shear Heel function Pack Go function Do not machine wash Machine wash at 70 C Machine wash at 95 C Chlorine Tumble dry Do not tumble dry Drip dry Do not iron Do not dry clean Wipe clean Class II equipment Dou ble insulated Indicated on the power adaptor Type BF IP class Enclosure class Temperature range The mattress should be used with the user lying lengthways C...

Страница 17: ...scription Size CC 901410 K Head cell PU 80x15 cm CC 901420 K Heel cell PU 80x13 cm CC 901412 K Head cell PU 85x15 cm CC 901422 K Heel cell PU 85x13 cm CC 901414 K Head cell PU 90x15 cm CC 901424 K Heel cell PU 90x13 cm CC 901416 K Head cell PU 105x15 cm CC 901426 K Heel cell PU 105x13 cm CC 901418 K Head cell PU 120x15 cm CC 901428 K Heel cell PU 120x13 cm Item no Description Size CC 901514 K Safe...

Страница 18: ...ce must always be performed by Care of Sweden or one of its authorised technicians Only use spare parts that have been approved by Care of Sweden Use of non approved spare parts invalidates the warranty For more information see the CuroCell UNO service manual Faults and defects regarded as warranty faults are fixed free of charge A warranty claim must be submitted to Care of Sweden before the prod...

Страница 19: ...ula tions It is recommended that this is performed by author ised service technicians or that the control unit is returned to Care of Sweden for scrapping Mattress A used CuroCell mattress should be taken to a recycling centre The product is sorted as combustible waste Care of Sweden complies with its manufacturer responsibility by being affiliated to the Swedish Packaging and Newspaper Collection...

Страница 20: ...ernet www careofsweden com Address Care of Sweden AB P O Box 146 SE 514 23 Tranemo SWEDEN Address for visitors Fabriksgatan 5A SE 514 33 Tranemo SWEDEN Cargo address Byns väg 4A SE 514 33 Tranemo Distributed by Care of Sweden and CuroCell UNO are trademarks of Care of Sweden AB Care of Sweden AB 2021 ...

Отзывы: