background image

WARNINGS AND SAFETY PRECAUTIONS

Read all instructions before use or repair

WARNING! To minimize the risk of fire, personal injury and equipment/property damage adhere to the following 

instructions:

1.  The product must only be installed and used for its intended purpose according to the instructions in this manual and/   

 

or other documentation from Care of Sweden. The product must not be combined, assembled or repaired with parts    

 

(e.g. control unit and mattress), accessories or spare parts other than those described in this manual or other 

 

documentation from Care of Sweden. The product must not be modified in any way.

2.  The product must be placed and used so that it does not become trapped or damaged. Note: Be particularly aware of    

 

trapping damage when using side rails. Always make sure that the mattress is the correct size for the bed.

3.  Regularly check product functionality by performing a hand check (see section 3.3).
4.  When the product is used for individuals needing special supervision, such as children, continuous monitoring is 

 required.

5.  The mattress is protected by a hygiene cover; avoid using multiple hygiene covers as this can affect the vapor 

 

permeability of the mattress.

6.  The hygiene cover does not allow liquid or air to penetrate, but is vapour permeable. Make sure that the user is 

 

positioned correctly to avoid the risk of suffocation.

7.  Be careful with sharp objects to prevent damage to the hygiene cover.
8.  Do not open the control unit housing – risk for electric shock. Servicing and maintenance must be performed by 

 

Care of Sweden or one of its authorized service technicians.

9.  Route the power cable to the control unit carefully to avoid tripping. Also make sure that the user is lying correctly on    

 

the mattress according to the instructions and use a cable holder if possible.

10. To avoid the risk of strangulation, make sure that the cable and tubes are routed to prevent someone getting caught up  

 

in them.

Note: The responsible healthcare provider must inform the user/operator of the following:

11. Do not use the product in bathrooms or other area where there is a risk of the control unit coming into contact with    

 

water or other liquids. Except for specified cleaning, never handle a product that has come into contact with 

 

water/liquid. Pull the plug out of the electrical socket immediately and send the product to an authorized service 

 

technician for servicing

12. Do not use close to or in contact with fire sources/hot surfaces, such as fire, burning cigarettes, hot lamps, heating fans   

 

or heating stoves/open fires as this could damage the product.

13. Do not store or use the product in direct sunlight. The product may be damaged by the elevated temperature and 

 

UV light.

 

14. Strong magnetic fields or wireless communication equipment (e.g. wireless home network products, mobile phones,    

 

walkie‐talkies, cordless phones and their base stations, radio transmitters, etc.) may affect the product’s functionality    

 

and should be kept at a distance of at least 1 meters from the control unit.

15. Never use the product if the power cable, plug of the control unit or power supply housing is defective, if the control    

 

unit housing is damaged, or if it is not functioning properly. Contact an authorized service technician for examination    

 

and repair.

16. Never connect anything other than the Care of Sweden supplied Power supply to the control unit power cable 

 connector.

 

17.  If the hygiene cover is equipped with side handles, these are intended for managing or relocating the mattress. 

 

Do not use the handles to lift the mattress with an user lying on it. All other use takes place under your own liability and  

 

is not covered by the product warranty.

18. To prevent the Power supply from being pulled out, exercise caution when there are children and pets in the 

 

environment around the equipment.

19. Use of this product adjacent to or stacked with other equipment should be avoided because it could result in improper 

 

operation. If such use is necessary, this product and the other equipment should be observed to verify that they are    

 

operating normally.

 

20. In order to minimize the risk of wounds occurring on the feet, make sure that the user doesn’t come into contact with    

 

the hangers of the control unit.

3

Содержание CuroCell UNO

Страница 1: ...Instructions for use 2021 06 09 CuroCell UNO Automatic air mattress systems Instructions for use item number 95 001401 EN0000 ...

Страница 2: ... 2 Product description 5 3 Assembly installation 8 4 Operation 8 5 Reuse cleaning and reconditioning 12 6 Storage 12 7 Maintenance 13 8 Troubleshooting 13 9 Technical description 14 10 Spare parts 17 11 Options and accessories 18 12 Warranty 18 13 Other information 19 2 ...

Страница 3: ...the following 11 Do not use the product in bathrooms or other area where there is a risk of the control unit coming into contact with water or other liquids Except for specified cleaning never handle a product that has come into contact with water liquid Pull the plug out of the electrical socket immediately and send the product to an authorized service technician for servicing 12 Do not use close...

Страница 4: ... diagnosis and general condition The mattress should only be used lying in the lengthwise direction on the mat tress with the feet at the end marked with the feet symbol No training or special skills are required for the use of Curo Cell A4 It is designed to be easy to handle and understand reading the quick guide and the instructions for use before use is recommended Prescription shall be made by...

Страница 5: ... and mattress 1 Mattress 2 CPR Quick deflation 3 Tube connection control unit 4 Control unit 1 2 3 4 Mattress Hygienic cover top part Cell holder integrated in inner cover Cells Safety mattress and tubing set Inner cover Cover bottom part 5 ...

Страница 6: ... cover It is easy to handle and maintain and conforms to the stringent requirements in respect of cleaning and hygiene The hygiene cover is vapour permeable i e the vapour is transported away thus reducing the risk of skin maceration Stone Cover A removable cover with a drip proof zip fastener on four sides and welded seams 2 2 Control unit 1 Tube CPR connection 4 Power switch On Off 1 0 8 Power a...

Страница 7: ...led low pressure 2 3 4 Maximum pressure Caring mode 4 With this function the entire mattress is inflated and provides fixed support This function automatically switches to the preced ing setting after approx 20 minutes This function should be used when nursing the user shifting the user s position or moving the user in or out of bed 2 3 5 Screen lock 5 Press the Screen Lock button to lock or unloc...

Страница 8: ... is found contact Care of Sweden or your local distributor before using the product NOTE Equipment with an electrical safety class of II must always be used for home care and in nursing homes see section 9 3 Labelling In hospitals and equivalent sites equipment with an electrical safety class of I or II may be used depending on whether or not the socket is earthed See section 9 1 Technical specifi...

Страница 9: ...re that the mattress is working properly This must be performed regularly we recommend once per shift or every eight hours Open the cover and insert a hand between the upper cells beneath the user s pelvic region centre of the mattress Check to ensure that there is a gap to the underlying mattress so that the user does not bottom out We recommend a gap of one or two finger widths If the gap is too...

Страница 10: ... control unit Notifications Notifications are shown with a warning triangle on the main menu with a dedicated display image or simply with a sound and flashing orange light in the function button on the front Notifications are silenced and acknowledged disabled using the function button on the front In the event of problems with Notifications see section 8 for troubleshooting Silent notifications ...

Страница 11: ...power adaptor can be removed from the wall socket and the control unit left over the end of the bed during transport For optimum comfort Care of Sweden recommends first placing the control unit in Constant low pressure mode and waiting for 5 minutes before disconnecting The mattress will retain air for at least eight hours If the control unit has a backup battery the control unit will automaticall...

Страница 12: ...ch as alcohol with or without tenside surfactants Isopropanol 70 oxidizing solutions chlorine max 1 or hydrogen peroxide max 1 5 5 2 Reconditioning CONTROL UNIT See section 5 1 for instructions MATTRESS 1 Disconnect the tube connector from the control unit and remove the air from the mattress 2 Remove the covers For more information of cleaning the covers see section 5 1 3 Move the mattress to a c...

Страница 13: ... user is bottoming out The mattress moves around Some cells have less air The control unit makes a noise vibrations can be felt Solution Check that the power adaptor has been connected to the mains supply Check that the LED on the power adaptor is showing green Check that the power adaptor has not become loose both the main supply and the control unit Check that the switch is set to 1 On If the co...

Страница 14: ...0 C Storage 20 C 45 C Transport 25 C 70 C Operation 15 93 non condensing Storage 93 non condensing 700 hPa 1060 hPa 10 x Ni MH 1 2 V 3500 mAh Class I Class II Type BF IP42 The device is not intended for use with flammable anaesthetic gases Mattress SPECIFICATION CuroCell UNO 80 85 90 105 120 x 200 210 220 x 20 cm 11 kg 80x200 cm Pulsating and Alternating Ungrounded AC outlet electrical safety clas...

Страница 15: ... key THE MATTRESS The mattress cover and inner cover are marked as shown below example For an explanation see section 9 4 Symbol key WARNING To avoid the risk of electric shock equipment marked as being in electrical safety class I must be connected to an earthed socket only Double insulation symbol denotes electrical safety class II may be connected to an unearthed socket Earth symbol denotes ele...

Страница 16: ...rectly on the bed base Counteracts shear Heel function Pack Go function Do not machine wash Machine wash at 70 C Machine wash at 95 C Chlorine Tumble dry Do not tumble dry Drip dry Do not iron Do not dry clean Wipe clean Class II equipment Dou ble insulated Indicated on the power adaptor Type BF IP class Enclosure class Temperature range The mattress should be used with the user lying lengthways C...

Страница 17: ...scription Size CC 901410 K Head cell PU 80x15 cm CC 901420 K Heel cell PU 80x13 cm CC 901412 K Head cell PU 85x15 cm CC 901422 K Heel cell PU 85x13 cm CC 901414 K Head cell PU 90x15 cm CC 901424 K Heel cell PU 90x13 cm CC 901416 K Head cell PU 105x15 cm CC 901426 K Heel cell PU 105x13 cm CC 901418 K Head cell PU 120x15 cm CC 901428 K Heel cell PU 120x13 cm Item no Description Size CC 901514 K Safe...

Страница 18: ...ce must always be performed by Care of Sweden or one of its authorised technicians Only use spare parts that have been approved by Care of Sweden Use of non approved spare parts invalidates the warranty For more information see the CuroCell UNO service manual Faults and defects regarded as warranty faults are fixed free of charge A warranty claim must be submitted to Care of Sweden before the prod...

Страница 19: ...ula tions It is recommended that this is performed by author ised service technicians or that the control unit is returned to Care of Sweden for scrapping Mattress A used CuroCell mattress should be taken to a recycling centre The product is sorted as combustible waste Care of Sweden complies with its manufacturer responsibility by being affiliated to the Swedish Packaging and Newspaper Collection...

Страница 20: ...ernet www careofsweden com Address Care of Sweden AB P O Box 146 SE 514 23 Tranemo SWEDEN Address for visitors Fabriksgatan 5A SE 514 33 Tranemo SWEDEN Cargo address Byns väg 4A SE 514 33 Tranemo Distributed by Care of Sweden and CuroCell UNO are trademarks of Care of Sweden AB Care of Sweden AB 2021 ...

Отзывы: