background image

F

G

H

I

J

Low power 

(during battery 

operation with 

battery backup)

Switching to 

battery operation 

start‐up

Charging battery

Battery operation

Restart

In the event 

of low power 

during battery 

operation, 

press the Func‐

tion button and 

the software 

will return to 

the main menu but with the ‘low battery’ 

symbol in the lower right‐hand corner lit 

(7). See section 2.3.
The program switches to Constant low 

pressure mode and no changes can be 

made before the control unit is connected 

to the mains power.
Connect the power adaptor to a power 

outlet.

When starting 

up the system 

on the backup 

battery, a 

confirmation 

is required by 

acknowledg‐

ing that the system should start and the 

images disappear.
Audio notification is also given in this case.
Connect the power adaptor to the power 

outlet for normal operation or acknowl‐

edge using the function button to operate 

the control unit by battery.

Shows that the 

mains power 

is connected 

and that the 

battery is 

charging.
Goes off automatically after 5 sec.

Shows that the 

control unit 

has switched 

over to battery 

operation.
Goes off auto‐

matically after 

5 sec. Connect 

the power adaptor to the power outlet for 

normal operation or let the control unit 

operate by battery.

When a restart 

is required, the 

control unit 

signals using a 

sound and the 

orange func‐

tion button.
Acknowledge using the function button 

and turn off the power by turning the On/

Off switch at the side of the control unit to 

0 (Off). Wait for approx. 10 seconds, then 

turn the control unit on again immediately 

and turn the On/Off switch to 1 (On).

4.7 Transport function

If the user needs to be moved in bed, the power adaptor 

can be removed from the wall socket and the control unit 

left over the end of the bed during transport. For optimum 

comfort, Care of Sweden recommends first placing the 

control unit in Constant low pressure mode and waiting for 5 

minutes before disconnecting. The mattress will retain air for 

at least eight hours.
If the control unit has a backup battery, the control unit 

will automatically switch to this when the mains cable is 

removed. The control unit will automatically switch to 

Constant low pressure mode and warn in good time before 

the battery is completely exhausted. The battery lasts for at 

least five hours.
We recommend using this function for short periods only.

4.8 Pack&Go®-function (deflation)

After use, the product can be packed as follows:

.

 Ensure that no‐one is lying on the mattress.

.

 On the main menu, press         in the top right hand corner.

.

 This takes you to the Information menu from which the 

Deflation function can be activated       .

.

 Confirm by pressing       .

The function button will flash orange during deflation. The 

mattress will empty of air and be ready to be simply folded 

together within 20 minutes. The control unit gives an audio 

signal once deflation is complete.
Carefully fold the mattress together, place the control unit

between the folds of the mattress and place in a transporta‐

tion bag (accessory) or equivalent for protective storage. En‐

sure that the power adaptor is packed away. After Pack&Go 

has been completed, the system returns to factory settings.
For any queries about installation, use or maintenance, 

please contact Care of Sweden or your local distributor.

4.9 Restart

If a restart is required, set the On/Off switch on the side of 

the control unit to position 0 (Off). Wait for approx. 10 

seconds and restart the control unit.
The control unit will now start up. See section 4.1.

11

Содержание CuroCell UNO

Страница 1: ...Instructions for use 2021 06 09 CuroCell UNO Automatic air mattress systems Instructions for use item number 95 001401 EN0000 ...

Страница 2: ... 2 Product description 5 3 Assembly installation 8 4 Operation 8 5 Reuse cleaning and reconditioning 12 6 Storage 12 7 Maintenance 13 8 Troubleshooting 13 9 Technical description 14 10 Spare parts 17 11 Options and accessories 18 12 Warranty 18 13 Other information 19 2 ...

Страница 3: ...the following 11 Do not use the product in bathrooms or other area where there is a risk of the control unit coming into contact with water or other liquids Except for specified cleaning never handle a product that has come into contact with water liquid Pull the plug out of the electrical socket immediately and send the product to an authorized service technician for servicing 12 Do not use close...

Страница 4: ... diagnosis and general condition The mattress should only be used lying in the lengthwise direction on the mat tress with the feet at the end marked with the feet symbol No training or special skills are required for the use of Curo Cell A4 It is designed to be easy to handle and understand reading the quick guide and the instructions for use before use is recommended Prescription shall be made by...

Страница 5: ... and mattress 1 Mattress 2 CPR Quick deflation 3 Tube connection control unit 4 Control unit 1 2 3 4 Mattress Hygienic cover top part Cell holder integrated in inner cover Cells Safety mattress and tubing set Inner cover Cover bottom part 5 ...

Страница 6: ... cover It is easy to handle and maintain and conforms to the stringent requirements in respect of cleaning and hygiene The hygiene cover is vapour permeable i e the vapour is transported away thus reducing the risk of skin maceration Stone Cover A removable cover with a drip proof zip fastener on four sides and welded seams 2 2 Control unit 1 Tube CPR connection 4 Power switch On Off 1 0 8 Power a...

Страница 7: ...led low pressure 2 3 4 Maximum pressure Caring mode 4 With this function the entire mattress is inflated and provides fixed support This function automatically switches to the preced ing setting after approx 20 minutes This function should be used when nursing the user shifting the user s position or moving the user in or out of bed 2 3 5 Screen lock 5 Press the Screen Lock button to lock or unloc...

Страница 8: ... is found contact Care of Sweden or your local distributor before using the product NOTE Equipment with an electrical safety class of II must always be used for home care and in nursing homes see section 9 3 Labelling In hospitals and equivalent sites equipment with an electrical safety class of I or II may be used depending on whether or not the socket is earthed See section 9 1 Technical specifi...

Страница 9: ...re that the mattress is working properly This must be performed regularly we recommend once per shift or every eight hours Open the cover and insert a hand between the upper cells beneath the user s pelvic region centre of the mattress Check to ensure that there is a gap to the underlying mattress so that the user does not bottom out We recommend a gap of one or two finger widths If the gap is too...

Страница 10: ... control unit Notifications Notifications are shown with a warning triangle on the main menu with a dedicated display image or simply with a sound and flashing orange light in the function button on the front Notifications are silenced and acknowledged disabled using the function button on the front In the event of problems with Notifications see section 8 for troubleshooting Silent notifications ...

Страница 11: ...power adaptor can be removed from the wall socket and the control unit left over the end of the bed during transport For optimum comfort Care of Sweden recommends first placing the control unit in Constant low pressure mode and waiting for 5 minutes before disconnecting The mattress will retain air for at least eight hours If the control unit has a backup battery the control unit will automaticall...

Страница 12: ...ch as alcohol with or without tenside surfactants Isopropanol 70 oxidizing solutions chlorine max 1 or hydrogen peroxide max 1 5 5 2 Reconditioning CONTROL UNIT See section 5 1 for instructions MATTRESS 1 Disconnect the tube connector from the control unit and remove the air from the mattress 2 Remove the covers For more information of cleaning the covers see section 5 1 3 Move the mattress to a c...

Страница 13: ... user is bottoming out The mattress moves around Some cells have less air The control unit makes a noise vibrations can be felt Solution Check that the power adaptor has been connected to the mains supply Check that the LED on the power adaptor is showing green Check that the power adaptor has not become loose both the main supply and the control unit Check that the switch is set to 1 On If the co...

Страница 14: ...0 C Storage 20 C 45 C Transport 25 C 70 C Operation 15 93 non condensing Storage 93 non condensing 700 hPa 1060 hPa 10 x Ni MH 1 2 V 3500 mAh Class I Class II Type BF IP42 The device is not intended for use with flammable anaesthetic gases Mattress SPECIFICATION CuroCell UNO 80 85 90 105 120 x 200 210 220 x 20 cm 11 kg 80x200 cm Pulsating and Alternating Ungrounded AC outlet electrical safety clas...

Страница 15: ... key THE MATTRESS The mattress cover and inner cover are marked as shown below example For an explanation see section 9 4 Symbol key WARNING To avoid the risk of electric shock equipment marked as being in electrical safety class I must be connected to an earthed socket only Double insulation symbol denotes electrical safety class II may be connected to an unearthed socket Earth symbol denotes ele...

Страница 16: ...rectly on the bed base Counteracts shear Heel function Pack Go function Do not machine wash Machine wash at 70 C Machine wash at 95 C Chlorine Tumble dry Do not tumble dry Drip dry Do not iron Do not dry clean Wipe clean Class II equipment Dou ble insulated Indicated on the power adaptor Type BF IP class Enclosure class Temperature range The mattress should be used with the user lying lengthways C...

Страница 17: ...scription Size CC 901410 K Head cell PU 80x15 cm CC 901420 K Heel cell PU 80x13 cm CC 901412 K Head cell PU 85x15 cm CC 901422 K Heel cell PU 85x13 cm CC 901414 K Head cell PU 90x15 cm CC 901424 K Heel cell PU 90x13 cm CC 901416 K Head cell PU 105x15 cm CC 901426 K Heel cell PU 105x13 cm CC 901418 K Head cell PU 120x15 cm CC 901428 K Heel cell PU 120x13 cm Item no Description Size CC 901514 K Safe...

Страница 18: ...ce must always be performed by Care of Sweden or one of its authorised technicians Only use spare parts that have been approved by Care of Sweden Use of non approved spare parts invalidates the warranty For more information see the CuroCell UNO service manual Faults and defects regarded as warranty faults are fixed free of charge A warranty claim must be submitted to Care of Sweden before the prod...

Страница 19: ...ula tions It is recommended that this is performed by author ised service technicians or that the control unit is returned to Care of Sweden for scrapping Mattress A used CuroCell mattress should be taken to a recycling centre The product is sorted as combustible waste Care of Sweden complies with its manufacturer responsibility by being affiliated to the Swedish Packaging and Newspaper Collection...

Страница 20: ...ernet www careofsweden com Address Care of Sweden AB P O Box 146 SE 514 23 Tranemo SWEDEN Address for visitors Fabriksgatan 5A SE 514 33 Tranemo SWEDEN Cargo address Byns väg 4A SE 514 33 Tranemo Distributed by Care of Sweden and CuroCell UNO are trademarks of Care of Sweden AB Care of Sweden AB 2021 ...

Отзывы: