5. Reuse, cleaning and reconditioning

The product is reusable. When reusing, it is important to 

follow the instructions below for cleaning and 

reconditioning. Before using the product again for a new 

user, it is recommended to clean the mattress in accordance 

with the instructions below. Consult your hygiene manager 

or Care of Sweden for help and instructions if you are unsure.

Cleaning and reconditioning should otherwise be performed 

as required.

5.1 Cleaning

CONTROL UNIT

Wipe off the control unit with a damp cloth and mild 

cleaning solution for this purpose: such as alcohol 

with or without tenside surfactants, oxidizing 

solutions, chlorine max 1 % or hydrogen peroxide max 1,5 %.
If using a different agent, choose one that does not affect the

control unit.

MATTRESS COVER

Covers can be wiped off with a mild cleaning solu‐ 

tion for this purpose such as: alcohol with or with‐ 

out tenside surfactants, Isopropanol 70 %, oxidizing

solutions, chlorine max 1 % (max 10% for cover Stone) or 

hydrogen peroxide max 1,5 %.
Excrement and blood stains must be removed as soon as 

possible using cold water. Carefully follow local instructions 

and the instructions for the detergent.

Multi‐layer covers should be divided before washing. The 

parts should be washed with similar colours.

Chlorine and phenol‐based cleaners could adversely 

affect the PU surface and should be avoided. If chlo‐ 

rine is used, we recommend a mixture of max 1 % 

(max 10% for cover Stone).

INNER COVER

Clean the affected area with a mild cleaning solution for this 

purpose such as: alcohol with or without tenside surfactants, 

Isopropanol 70 %, oxidizing solutions, chlorine max 1 % or 

hydrogen peroxide max 1,5 %.

5.2 Reconditioning

CONTROL UNIT

See section 5.1 for instructions.

MATTRESS

1. Disconnect the tube connector from the control unit and 

remove the air from the mattress.
2. Remove the covers. For more information of cleaning the 

covers, see section 5.1.
3. Move the mattress to a clean area that is suitable for 

cleaning.
4. Wipe off the cells, all tubes and the CPR module with a 

mild cleaning solution for this purpose such as: alcohol with 

or without tenside surfactants, Isopropanol 70 %, oxidizing 

solutions, chlorine max 1 % or hydrogen peroxide max 1,5 %. 

Let it dry.
5. Wipe off the work surfaces using a suitable detergent and/ 

or disinfectant.
6. Put the mattress together. If cells for any reason have be‐ 

come loose from the tubes, these must be replaced accord‐ 

ing to the drawing in section 2.1.

6. Storage

It is advisable to store the mattress and control unit in the 

product bag (accessory). Handle the packaged product with 

care. Do not place any heavy objects on top of it. For addi‐

tional information about storage temperature, see section 9.

If the control unit has a backup battery (variant) this must be 

charged up every six months if the control unit is not in use.
The control unit should be stored in a dry environment

indoors and out of direct sunlight.

12

NOTE:

Check the hygiene cover, cells and hoses each time the prod‐

uct is cleaned. If damaged, it must be replaced or repaired. 

Also check the control unit, tube connectors and power cable 

during cleaning. Damaged parts must be replaced or repaired.
The air cells can be divided using a quick‐release connector, 

making it easier to clean the cells.

Содержание CuroCell UNO

Страница 1: ...Instructions for use 2021 06 09 CuroCell UNO Automatic air mattress systems Instructions for use item number 95 001401 EN0000 ...

Страница 2: ... 2 Product description 5 3 Assembly installation 8 4 Operation 8 5 Reuse cleaning and reconditioning 12 6 Storage 12 7 Maintenance 13 8 Troubleshooting 13 9 Technical description 14 10 Spare parts 17 11 Options and accessories 18 12 Warranty 18 13 Other information 19 2 ...

Страница 3: ...the following 11 Do not use the product in bathrooms or other area where there is a risk of the control unit coming into contact with water or other liquids Except for specified cleaning never handle a product that has come into contact with water liquid Pull the plug out of the electrical socket immediately and send the product to an authorized service technician for servicing 12 Do not use close...

Страница 4: ... diagnosis and general condition The mattress should only be used lying in the lengthwise direction on the mat tress with the feet at the end marked with the feet symbol No training or special skills are required for the use of Curo Cell A4 It is designed to be easy to handle and understand reading the quick guide and the instructions for use before use is recommended Prescription shall be made by...

Страница 5: ... and mattress 1 Mattress 2 CPR Quick deflation 3 Tube connection control unit 4 Control unit 1 2 3 4 Mattress Hygienic cover top part Cell holder integrated in inner cover Cells Safety mattress and tubing set Inner cover Cover bottom part 5 ...

Страница 6: ... cover It is easy to handle and maintain and conforms to the stringent requirements in respect of cleaning and hygiene The hygiene cover is vapour permeable i e the vapour is transported away thus reducing the risk of skin maceration Stone Cover A removable cover with a drip proof zip fastener on four sides and welded seams 2 2 Control unit 1 Tube CPR connection 4 Power switch On Off 1 0 8 Power a...

Страница 7: ...led low pressure 2 3 4 Maximum pressure Caring mode 4 With this function the entire mattress is inflated and provides fixed support This function automatically switches to the preced ing setting after approx 20 minutes This function should be used when nursing the user shifting the user s position or moving the user in or out of bed 2 3 5 Screen lock 5 Press the Screen Lock button to lock or unloc...

Страница 8: ... is found contact Care of Sweden or your local distributor before using the product NOTE Equipment with an electrical safety class of II must always be used for home care and in nursing homes see section 9 3 Labelling In hospitals and equivalent sites equipment with an electrical safety class of I or II may be used depending on whether or not the socket is earthed See section 9 1 Technical specifi...

Страница 9: ...re that the mattress is working properly This must be performed regularly we recommend once per shift or every eight hours Open the cover and insert a hand between the upper cells beneath the user s pelvic region centre of the mattress Check to ensure that there is a gap to the underlying mattress so that the user does not bottom out We recommend a gap of one or two finger widths If the gap is too...

Страница 10: ... control unit Notifications Notifications are shown with a warning triangle on the main menu with a dedicated display image or simply with a sound and flashing orange light in the function button on the front Notifications are silenced and acknowledged disabled using the function button on the front In the event of problems with Notifications see section 8 for troubleshooting Silent notifications ...

Страница 11: ...power adaptor can be removed from the wall socket and the control unit left over the end of the bed during transport For optimum comfort Care of Sweden recommends first placing the control unit in Constant low pressure mode and waiting for 5 minutes before disconnecting The mattress will retain air for at least eight hours If the control unit has a backup battery the control unit will automaticall...

Страница 12: ...ch as alcohol with or without tenside surfactants Isopropanol 70 oxidizing solutions chlorine max 1 or hydrogen peroxide max 1 5 5 2 Reconditioning CONTROL UNIT See section 5 1 for instructions MATTRESS 1 Disconnect the tube connector from the control unit and remove the air from the mattress 2 Remove the covers For more information of cleaning the covers see section 5 1 3 Move the mattress to a c...

Страница 13: ... user is bottoming out The mattress moves around Some cells have less air The control unit makes a noise vibrations can be felt Solution Check that the power adaptor has been connected to the mains supply Check that the LED on the power adaptor is showing green Check that the power adaptor has not become loose both the main supply and the control unit Check that the switch is set to 1 On If the co...

Страница 14: ...0 C Storage 20 C 45 C Transport 25 C 70 C Operation 15 93 non condensing Storage 93 non condensing 700 hPa 1060 hPa 10 x Ni MH 1 2 V 3500 mAh Class I Class II Type BF IP42 The device is not intended for use with flammable anaesthetic gases Mattress SPECIFICATION CuroCell UNO 80 85 90 105 120 x 200 210 220 x 20 cm 11 kg 80x200 cm Pulsating and Alternating Ungrounded AC outlet electrical safety clas...

Страница 15: ... key THE MATTRESS The mattress cover and inner cover are marked as shown below example For an explanation see section 9 4 Symbol key WARNING To avoid the risk of electric shock equipment marked as being in electrical safety class I must be connected to an earthed socket only Double insulation symbol denotes electrical safety class II may be connected to an unearthed socket Earth symbol denotes ele...

Страница 16: ...rectly on the bed base Counteracts shear Heel function Pack Go function Do not machine wash Machine wash at 70 C Machine wash at 95 C Chlorine Tumble dry Do not tumble dry Drip dry Do not iron Do not dry clean Wipe clean Class II equipment Dou ble insulated Indicated on the power adaptor Type BF IP class Enclosure class Temperature range The mattress should be used with the user lying lengthways C...

Страница 17: ...scription Size CC 901410 K Head cell PU 80x15 cm CC 901420 K Heel cell PU 80x13 cm CC 901412 K Head cell PU 85x15 cm CC 901422 K Heel cell PU 85x13 cm CC 901414 K Head cell PU 90x15 cm CC 901424 K Heel cell PU 90x13 cm CC 901416 K Head cell PU 105x15 cm CC 901426 K Heel cell PU 105x13 cm CC 901418 K Head cell PU 120x15 cm CC 901428 K Heel cell PU 120x13 cm Item no Description Size CC 901514 K Safe...

Страница 18: ...ce must always be performed by Care of Sweden or one of its authorised technicians Only use spare parts that have been approved by Care of Sweden Use of non approved spare parts invalidates the warranty For more information see the CuroCell UNO service manual Faults and defects regarded as warranty faults are fixed free of charge A warranty claim must be submitted to Care of Sweden before the prod...

Страница 19: ...ula tions It is recommended that this is performed by author ised service technicians or that the control unit is returned to Care of Sweden for scrapping Mattress A used CuroCell mattress should be taken to a recycling centre The product is sorted as combustible waste Care of Sweden complies with its manufacturer responsibility by being affiliated to the Swedish Packaging and Newspaper Collection...

Страница 20: ...ernet www careofsweden com Address Care of Sweden AB P O Box 146 SE 514 23 Tranemo SWEDEN Address for visitors Fabriksgatan 5A SE 514 33 Tranemo SWEDEN Cargo address Byns väg 4A SE 514 33 Tranemo Distributed by Care of Sweden and CuroCell UNO are trademarks of Care of Sweden AB Care of Sweden AB 2021 ...