17
9.1
Technical specification
NB:
Care of Sweden reserves the right to modify the product specification at any time.
CONTROL UNIT SPECIFICATION
Model
CuroCell® IQ Cirrus
Input voltage
100-240 V / 50-60 Hz
Power consumption
1.5 A
Mode of operation
Pulsating
Duty cycle
10 min
Power supply
Ungrounded AC outlet, electrical safety
class II
Use only power supply with P/N WR9QE1500LRPCIMG3138
Dimensions (L x W x H)
11 cm x 30 cm x 20 cm
Weight
2.9 kg
Sound level, control unit,
max:
17 dBA (at operator position), 16,5 dBA (at head end). Specifies the
highest value in accordance with EN ISO 11201:2010 when placing
the control unit on the foot end.
Environmental
Temperature
Operation: +5 to 40°C
Storage: -25 to 70°C
Transport: -25 to 70 °C
Humidity
Operation: 15 % – 93 % non-condensing
Storage: < 93 % non-condensing
Atmospheric
700 hPa – 1060 hPa
Electrical classification
Class II, Type BF
IP classification
IP42
Degree of safety in
presence of inflammable
anesthetics:
The device is not intended for use with flammable anesthetic gases
Applied part
Mattress
MATTRESS SPECIFICATION
Model
Dimensions (WxLxH)
Weight
CuroCell® Cirrus
80/85/90/100/105/120x200x21 cm
18 kg (width 85 cm)
9.2
Standards
The system is tested and approved according to the following European standards where applicable requirements are met.
IEC 60601-1
EN ISO 10993
ISO 3746
IEC 60601-1-2
EN 12182
ISO 11201
IEC 60601-1-11
IEC 60601-1-6
IEC 62304
EN 597-1
EN 597-2
EN ISO 14971
EN 14126
9. Technical description