3.6.1 Table of notifications 

 

Information about each notification is shown in the 

notification table: 

 

Notifications from 1-10 are both audible and visual.          

The error code will be displayed until the error has been 

rectified. If the mute button is pressed the audible 

warning will crease for a period of 5 minutes and will 

return until the error has been rectified. 

 

The notifications 11-13 have no audible alarm. The error 

code is shown until the system is restarted. 

 
 

Notification 

Description and 

troubleshooting 

 

 

 

High temperature. Valves and 

compressors are turned off. 

If the control unit is in direct 

sunlight, relocate it. 

Otherwise, contact the 

support. 

 

 

 

 

     

Default settings are not 

completed. Contact the 

support. 

 

 

Incorrect input voltage. Make 

sure that the correct power 

supply is used, otherwise 

contact the support. 

 

 

Low pressure. Secure the 

CPR, mattress, air tubes and 

air filter. If the problem 

remains, contact the support. 

 

 

 

Automatic setting failure. The 

correct pressure has not been 

reached within the time limit. 

If the problem remains, 

contact the support.  

 

 

 

 

Leakage under the automatic 

setting period. The mattress 

is leaking too much for the 

weighing to be accomplished. 

Control the mattress and the 

air connections. If the 

problem remains, contact the 

support. 

 

 

 

High pressure. The pressure 

cannot be reduced to the 

desired value within the time 

limit. Contact the support.  

 

 

 

The automatic setting has 

been restarted too many 

times during the automatic 

setting period. Contact the 

support. 

 

 

The mattress control 

parameters have not been 

read. Connect the CPR or 

contact the support. 

 

  

 

 

10 

 

The mattress control 

parameters have been 

changed during the use. 

Restart the system. If the 

problem remains, contact the 

support. 

 

 

11 

 

Leakage in one of the cell 

sections. Secure the CPR, 

mattress and connection 

tubes. If the problem 

remains, contact the support.  

 

 

12 

 

Leakage in one of the cell 

sections. Secure the CPR, 

mattress and connection 

tubes. If the problem 

remains, contact the support. 

 

 

13 

 

Leakage in one of the cell 

sections. Secure the CPR, 

mattress and connection 

tubes. If the problem 

remains, contact the support. 

 

 

3.6.2 Maximum pressure notification 

 

When the function Maximum pressure has 

been used for a long time, the Maximum 

pressure diode will blink. If the use is 

intentional, ignore the notification. 

 

 

3.7 Transport function 

If the user needs to be moved in bed, there are two ways 

to perform the transportation: 

 

Unplug the CPR connection, close the lid, place the CPR 

connection at the end of the bed and remove the 

control unit from the bed. The mattress will retain air 

for at least 12 hours. 

or 

 

Remove the power supply from the wall socket and 

leave the control unit hanging on the bed during 

transport. The mattress will retain air for at least 12

 

hours. 

We recommend using this function for short periods only. 
 
 
 
 
 
 
 
 
 
 
 

Содержание CuroCell IQ Cirrus

Страница 1: ...Instructions for use 2020 08 14 CuroCell IQ Cirrus Automatic active airflow controlled mattress system Instructions for use item number 95 001450 EN0000 ...

Страница 2: ...4 2 Assembly installation 6 3 Operation 7 4 Product description 12 5 Reuse cleaning and reconditioning 14 6 Storage 14 7 Maintenance 15 8 Troubleshooting 16 9 Technical description 17 10 Spare parts 21 11 Options and accessories 22 12 Warranty 22 13 Other information 23 ...

Страница 3: ... other area where there is a risk of the control unit coming into contact with water or other liquids Except for specified cleaning never handle a product that has come into contact with water liquid Pull the plug out of the electrical socket immediately and send the product to an authorized service technician for servicing 12 Do not use close to or in contact with fire sources hot surfaces such a...

Страница 4: ...ponsible personnel depending on the user s status diagnosis and general condition The mattress should only be used lying in the lengthwise direction on the mat tress with the feet at the end marked with the feet symbol No training or special skills are required for the use of CuroCell IQ Cirrus It is designed to be easy to handle and understand reading the quick guide and the instructions for use ...

Страница 5: ...head in the bed when pulling down Pulling in the stairs Start with pulling an empty mattress down the stairs in order to practice the technique When training with a person on the mattress a string or rope can be fasten in the handles at the head end Another person can then walk behind the mattress and help out if problems should occur It is advisable to practice in the same stairs where an evacuat...

Страница 6: ...mmed when raising or lowering the bed 4 Open the lid on the air tube connector marked CPR and connect it to the side of the control unit 5 A click is heard and felt when correctly connected Secure that both sides of the connection are closed 6 Check that the switch on the side of the control unit is set to 0 off Plug the power supply into an approved electrical socket 100 240 V 7 Check that the po...

Страница 7: ...nflated 3 The diode over the checkmark will blink during the inflation Depending on its size it takes approximately 20 30 minutes for the mattress to inflate fully 4 When the green diode over the checkmark is constant the system has adapted the settings to the user s individual needs and the system can be taken into use This takes approximately 20 30 minutes 3 Operation NOTE During the automatic s...

Страница 8: ...ons Button Function Mute the information signal Pack Go Automatic deflation of the system Maximum pressure caring mode Information signal Incorrect connection of the air connector CPR System is ready to use Notification diodes ...

Страница 9: ...roperly This must be performed regularly we recommend once per shift or every eight hours Open the cover and insert a hand between the upper cells beneath the user s sacrum center of the mattress Check to ensure that there is a gap to the underlying surface so that the user does not bottom out If you can feel the user s sacrum resting in the palm of your hand the gap is too small Make a new automa...

Страница 10: ...ot be reduced to the desired value within the time limit Contact the support 8 The automatic setting has been restarted too many times during the automatic setting period Contact the support 9 The mattress control parameters have not been read Connect the CPR or contact the support 10 The mattress control parameters have been changed during the use Restart the system If the problem remains contact...

Страница 11: ...nutes The control unit gives an audio signal once deflation is complete Carefully fold the mattress together place the control unit between the folds of the mattress and place the system in a transportation bag accessory or equivalent for protective storage Ensure that the power supply is packed complete 3 9 Restart If a restart is required set the On Off switch on the side of the control unit to ...

Страница 12: ...nd hygiene The hygiene cover is vapour permeable i e the vapour is transported away thus reducing the risk of skin maceration The following covers are available for the CuroCell Cirrus mattress Cover Stone Evac A removable cover with a drip proof zip fasterner and welded seams Divisible top and bottom part The Evac cover is equipped with three handles on the head and foot end On the sides there ar...

Страница 13: ...13 CuroCell IQ Cirrus 1 Mattress 2 Tube connection CPR quick deflation 3 Control unit 1 2 3 CuroCell IQ Cirrus with cover Stone Hygiene cover Cells Foam core Inner cover Tubing set Bottom part ...

Страница 14: ...1 or hydrogen peroxide max 1 5 5 2 Reconditioning CONTROL UNIT See section 5 1 for instructions MATTRESS 1 Disconnect the tube connector from the control unit and remove the air from the mattress 2 Remove the covers For more information of cleaning the covers see section 5 1 3 Move the mattress to a clean area that is suitable for cleaning 4 Wipe off the cells all tubes and the CPR module with a m...

Страница 15: ...be submitted to Care of Sweden be fore the product is returned After the product is used check that 1 The power cable and power supply are undamaged 2 The connecting tubes marked CPR on the side of the control unit are positioned correctly and not leaking 3 The hygiene cover is intact and the cover and cells are correctly assembled 4 No tubes or connectors are damaged or jammed Contact Care of Swe...

Страница 16: ...raps underneath two at the head end and two on each of the long sides Some cells have less air This is normal with a Pulsating or Alternating mode as the air supply switches between alternating cells for a predetermined cycle period one cycle 10 minutes The control unit makes a noise vibrations can be felt Check how the control unit is hung on the bed Resonance can occur vibrations are felt in par...

Страница 17: ... foot end Environmental Temperature Operation 5 to 40 C Storage 25 to 70 C Transport 25 to 70 C Humidity Operation 15 93 non condensing Storage 93 non condensing Atmospheric 700 hPa 1060 hPa Electrical classification Class II Type BF IP classification IP42 Degree of safety in presence of inflammable anesthetics The device is not intended for use with flammable anesthetic gases Applied part Mattres...

Страница 18: ...nit is marked as shown below example For an explanation see section 9 4 Explanation of symbols POWER SUPPLY The power supply is part of the control unit and is marked as shown below example For an explanation see section 9 4 Explanation of symbols ...

Страница 19: ...19 THE MATTRESS The mattress cover and inner cover are marked as shown below example For an explanation see section 9 4 Explanation of symbols ...

Страница 20: ...n Place on top of existing mattress Do not machine wash Machine wash at 70 C Machine wash at 95 C Chlorine Tumble dry Do not tumble dry Drip dry Do not iron Do not dry clean Wipe clean Do not rotate Do not turn around Fire requirements Minimum length Serial number Class II equipment Dou ble insulated Indicated on the power supply IP class Enclosure class The mattress should be Distributed by Medic...

Страница 21: ...0 100x200 cm CT150 100200 EN0000 CB150 100200 EN0000 CC150 100200 EN0000 100x210 cm CT150 100210 EN0000 CB150 100210 EN0000 CC150 100210 EN0000 100x220 cm CT150 100220 EN0000 CB150 100220 EN0000 CC150 100220 EN0000 105x200 cm CT150 105200 EN0000 CB150 105200 EN0000 CC150 105200 EN0000 105x210 cm CT150 105210 EN0000 CB150 105210 EN0000 CC150 105210 EN0000 105x220 cm CT150 105220 EN0000 CB150 105220...

Страница 22: ...10 EN0000 120x210 cm CI101 120220 EN0000 120x220 cm Sturdy transportation bag Item no Model and size 58 900905 Transport bag 100 120 cm 58 900906 Transport bag extra high 80 90 cm 58 900911 Transport bag extra high 100 120 cm 11 3 Other accessories To ensure the best level of comfort we recommend use of accessories from Care of Sweden since our products are designed to work well together If you re...

Страница 23: ...r parts of the control unit must not be disassembled and are sorted as electronic waste Mattress A used CuroCell mattress should be taken to a recycling center The product is sorted as combustible waste Care of Sweden complies with its manufacturer responsibility by being affiliated to the Swedish Packaging and Newspaper Collection Service FTI and the electrical and electronic product recycling se...

Страница 24: ...rnet www careofsweden com Address Care of Sweden AB P O Box 146 SE 514 23 Tranemo SWEDEN Address for visitors Fabriksgatan 5A SE 514 33 Tranemo SWEDEN Cargo address Byns väg 4A SE 514 33 Tranemo SWEDEN Distributed by Care of Sweden and CuroCell are trademarks of Care of Sweden AB Care of Sweden AB 2020 ...

Отзывы: