INSTALLATION, Cont. 

Ethernet Cable Installation 

NOTE:

  

The Ethernet cable must be routed through the provided special 

metallic gland connector (CPN 6610-1506) and the shield wire must be 
connected to this gland connector for grounding and to eliminate RFI.   

 

1.  Remove the acorn nuts securing the back panel to 

main housing. 

2.  Locate and remove an unused cable gland 

connector for the cable. 

3.  Install the provided special metallic gland connector. 

4.  Loosen and remove the metal gland connector nut 

and remove the plastic insert. 

5.  Route the Ethernet cable (CAT 5 or equivalent) 

through the nut and plastic insert and into the 
enclosure. 

6.  With the Ethernet cable routed into the enclosure, 

refer to the figure below and remove the cable outer 
insulating jacket exposing the internal wires.  Allow an adequate length of exposed wires to 
reach the 2XX-IP Ethernet port (without being tight) after it has been terminated. 

7.  Next, cut the shield wire so that it extends past the outer jacket approximately 3/4 inch. 

 
8.  Referring to the figure below, fold the shield wire back over the plastic insert and then insert 

the plastic insert (with the shield wire) into the gland connector. 

 

 

 3/4 inch

Remove outer jacket 

 Shield 
 Wire 

Содержание 2XX-IP

Страница 1: ...For 200 Series Indicators 205 210 212G 212GX and 225 INSTALLATION MANUAL 8200 M503 O1 Rev B 203 E Daugherty Webb City MO 64870 USA Printed in USA 10 14 Ph 417 673 4631 Fax 417 673 2153 http www cardin...

Страница 2: ...8200 M503 01 Rev B 2XX IP 2...

Страница 3: ...are designed to provide reasonable protection against such interference when operated in a commercial environment Operation of this equipment in a residential area may cause interference in which case...

Страница 4: ...nt panel to the rear housing 4 Separate the rear panel from the front housing on the 205 210 and 225 or the front panel from the rear housing on the 212G 212GX 5 Lift the panel from the main housing t...

Страница 5: ...P Option Card Cont 205 210 212G 212GX Main PCB NOTE 205 Main PCB shown The 210 212G 212GX Main PCB will have connector P2 REMOTE OUTPUTS and J7 REMOTE OUTPUT SRC jumper installed 225 Main PCB P15 OPTI...

Страница 6: ...nut and remove the plastic insert 5 Route the Ethernet cable CAT 5 or equivalent through the nut and plastic insert and into the enclosure 6 With the Ethernet cable routed into the enclosure refer to...

Страница 7: ...OLS Finger tighten only Insure any unused gland connectors are plugged 2 Make certain no cables or wires are exposed between the main housing and rear panel 3 Place the rear panel onto the main housin...

Страница 8: ...Output No Continuous Output If Cont4 YES Continuous Output is selected an additional prompt tYPE4 will be displayed If Cont4 no No Continuous Output is selected the indicator will operate in the Weig...

Страница 9: ...key then press the NEXT Navigation arrow until SETUP MENU 3 is displayed SETUP MENU 3 1 G N ACCUMS XXX 6 KEY LOCKOUT 2 BACK LITE X 7 LRPort COM3 3 PASSWORD 4 2XX IP 5 BADGE RDR Enter Selection 0 PREV...

Страница 10: ...e 2XX IP card can be selected Press the 7 key and then the ENTER key 2XX IP SETUP MENU SCALE XXX Allowable values are 0 through 5 NOTE Scale 1 is the default setting Also note 4 is the Scale Totalizer...

Страница 11: ...LED is not flashing then communications between the 2XX IP and the indicator has failed XPORT Status LED s LED s are provided on the edge of the XPORT connector for diagnostics Their status and meani...

Страница 12: ...8200 M503 01 Rev B 2XX IP 10...

Отзывы: