8200-M503-01 Rev-B
•
2XX-IP
2
INSTALLATION
Mounting the 2XX-IP Board
NOTE:
Should your indicator come with the 2XX-IP board already installed, the following
information describing the mounting of the board does not apply. Proceed to the RIO Link
Cable Installation section.
CAUTION! This board contains static sensitive components. Improper
handling can result in damage to or destruction of the components or board.
Such actual and/or consequential damage IS NOT covered under warranty.
1. Make sure the power to the indicator is OFF. Unplug
the AC power cord.
2. Loosen the gland connectors where the AC power cord
and any other cables enter the indicator.
3. Remove the rear panel on the 205/210 and 225 or the
front panel on the 212G/212GX.
•
On the 205/210, remove the 12 acorn nuts securing
the rear panel to the front housing.
•
On the 225, remove the 14 acorn nuts securing the
rear panel to the front housing.
•
On the 212G/212GX, remove the 12 mounting
screws securing the front panel to the rear housing.
4. Separate the rear panel from the front housing on the
205/210 and 225 or the front panel from the rear
housing on the 212G/212GX.
5. Lift the panel from the main housing, taking care not to
stretch the cable and wires between the panel and
main housing. Lay the panel on the workbench next to
the indicator.
6. Referring to the illustrations on page 4, locate the
threaded stand-off and the OPTION BOARD connector
on the indicator main board.
205/210/212G/212GX – The threaded stand-off is
below P3 and the OPTION BOARD connector is
P10.
225 – The threaded stand-off is near P9 and the
OPTION BOARD connector is P15.
7. To install the 2XX-IP card, carefully align the connector
P7 (pins on bottom side of 2XX-IP board) with
connector P10 on the 205/210/212G/212GX or P15 on
the 225 main board and then apply even downward
pressure to the end of the 2XX-IP board (with P7 pins
on bottom side).
8. Align the hole in the 2XX-IP board with the threaded stand-off on the indicator main board
and using the lock washer and screw supplied with the 2XX-IP board, secure the 2XX-IP to
the indicator main board.
P1 (Pins on trace
side of board)
Ethernet Port
GND