8200-M503-01 Rev-B 

 2XX-IP 

INSTALLATION 

Mounting the 2XX-IP Board 

NOTE:

  Should your indicator come with the 2XX-IP board already installed, the following 

information describing the mounting of the board does not apply.  Proceed to the RIO Link 
Cable Installation section. 
 

 

CAUTION!  This board contains static sensitive components.  Improper 
handling can result in damage to or destruction of the components or board.  
Such actual and/or consequential damage IS NOT covered under warranty. 

 
1.  Make sure the power to the indicator is OFF.  Unplug 

the AC power cord. 

2.  Loosen the gland connectors where the AC power cord 

and any other cables enter the indicator. 

3.  Remove the rear panel on the 205/210 and 225 or the 

front panel on the 212G/212GX. 

 

On the 205/210, remove the 12 acorn nuts securing 
the rear panel to the front housing. 

 

On the 225, remove the 14 acorn nuts securing the 
rear panel to the front housing. 

 

On the 212G/212GX, remove the 12 mounting 
screws securing the front panel to the rear housing. 

4.  Separate the rear panel from the front housing on the 

205/210 and 225 or the front panel from the rear 
housing on the 212G/212GX. 

5.  Lift the panel from the main housing, taking care not to 

stretch the cable and wires between the panel and 
main housing.  Lay the panel on the workbench next to 
the indicator. 

6.  Referring to the illustrations on page 4, locate the 

threaded stand-off and the OPTION BOARD connector 
on the indicator main board. 

  205/210/212G/212GX – The threaded stand-off is 

below P3 and the OPTION BOARD connector is 
P10. 

  225 – The threaded stand-off is near P9 and the 

OPTION BOARD connector is P15. 

7.  To install the 2XX-IP card, carefully align the connector 

P7 (pins on bottom side of 2XX-IP board) with 
connector P10 on the 205/210/212G/212GX or P15 on 
the 225 main board and then apply even downward 
pressure to the end of the 2XX-IP board (with P7 pins 
on bottom side). 

8.  Align the hole in the 2XX-IP board with the threaded stand-off on the indicator main board 

and using the lock washer and screw supplied with the 2XX-IP board, secure the 2XX-IP to 
the indicator main board. 

P1 (Pins on trace 
side of board) 

Ethernet Port

GND 

Содержание 2XX-IP

Страница 1: ...For 200 Series Indicators 205 210 212G 212GX and 225 INSTALLATION MANUAL 8200 M503 O1 Rev B 203 E Daugherty Webb City MO 64870 USA Printed in USA 10 14 Ph 417 673 4631 Fax 417 673 2153 http www cardin...

Страница 2: ...8200 M503 01 Rev B 2XX IP 2...

Страница 3: ...are designed to provide reasonable protection against such interference when operated in a commercial environment Operation of this equipment in a residential area may cause interference in which case...

Страница 4: ...nt panel to the rear housing 4 Separate the rear panel from the front housing on the 205 210 and 225 or the front panel from the rear housing on the 212G 212GX 5 Lift the panel from the main housing t...

Страница 5: ...P Option Card Cont 205 210 212G 212GX Main PCB NOTE 205 Main PCB shown The 210 212G 212GX Main PCB will have connector P2 REMOTE OUTPUTS and J7 REMOTE OUTPUT SRC jumper installed 225 Main PCB P15 OPTI...

Страница 6: ...nut and remove the plastic insert 5 Route the Ethernet cable CAT 5 or equivalent through the nut and plastic insert and into the enclosure 6 With the Ethernet cable routed into the enclosure refer to...

Страница 7: ...OLS Finger tighten only Insure any unused gland connectors are plugged 2 Make certain no cables or wires are exposed between the main housing and rear panel 3 Place the rear panel onto the main housin...

Страница 8: ...Output No Continuous Output If Cont4 YES Continuous Output is selected an additional prompt tYPE4 will be displayed If Cont4 no No Continuous Output is selected the indicator will operate in the Weig...

Страница 9: ...key then press the NEXT Navigation arrow until SETUP MENU 3 is displayed SETUP MENU 3 1 G N ACCUMS XXX 6 KEY LOCKOUT 2 BACK LITE X 7 LRPort COM3 3 PASSWORD 4 2XX IP 5 BADGE RDR Enter Selection 0 PREV...

Страница 10: ...e 2XX IP card can be selected Press the 7 key and then the ENTER key 2XX IP SETUP MENU SCALE XXX Allowable values are 0 through 5 NOTE Scale 1 is the default setting Also note 4 is the Scale Totalizer...

Страница 11: ...LED is not flashing then communications between the 2XX IP and the indicator has failed XPORT Status LED s LED s are provided on the edge of the XPORT connector for diagnostics Their status and meani...

Страница 12: ...8200 M503 01 Rev B 2XX IP 10...

Отзывы: