Getting Started
2-13
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
4.
Prime, as follows:
a. Orient pressure sensing disc with patient side up.
b. Depress and hold pressure sensing disc between
2
fingers.
A
Prime Set with S
y
ringe
>Press and Hold PRIME
Prime Volume
0
mL
CAUTION: Do
not prime with
disposable set
connected to
patient!
PRIME
EXIT
c. Press and hold
PRIME
soft key until fluid flows and
priming of syringe administration set is complete.
d. Release pressure sensing disc.
A
Prime Set with S
y
ringe
>Press and Hold PRIME
Prime Volume
2 mL
CAUTION: Do
not prime with
disposable set
connected to
patient!
PRIME
EXIT
5.
When priming is complete, release
PRIME
soft key.
5
10
15
20
½
Patient Side
Priming
(continued)
Preparing Syringe and Administration Set (
Syringe Module
)
(Continued)
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
Pressure Sensing Disc Installed
(Continued)
Содержание Alaris 8000 Series
Страница 8: ......
Страница 52: ......
Страница 54: ......
Страница 176: ......
Страница 242: ......
Страница 284: ......
Страница 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Страница 318: ......
Страница 338: ......
Страница 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Страница 340: ......
Страница 350: ......
Страница 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Страница 352: ......
Страница 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Страница 362: ......
Страница 377: ......
Страница 378: ......