3-58
Troubleshooting and Maintenance
Alaris
®
System Directions for Use
Patient Controlled Analgesia Module Section
Alarms, Errors, Messages
(Continued)
Alarms
Attach Dose Request Cord
Channel Disconnected
Lock Door
Occlusion
PCA Pause Alarm
Dose Request Cord detached from
device. Dose Request Cord
required for
PCA
only and
PCA
+
continuous infusion modes.
Module(s) disconnected while in
operation or have a communication
problem.
Door unlocked during infusion.
System will not infuse with door
unlocked.
Increased back pressure sensed
while infusing. Infusion stops on
affected module.
PCA infusion has paused due to a
decline in respiratory status.
Reattach Dose Request Cord and
press
RESTART
key.
To silence alarm and clear
message from screen, press
CONFIRM
soft key. Reattach
module, if desired, ensuring it is
securely “clicked” into place at
Channel Release Latch. If alarm is
still present, replace module with
an operational instrument.
Lock door and press
RESTART
key.
Clear occlusion. Press
RESTART
key, or press
CHANNEL SELECT
key and then
START
soft key.
Assess patient status per hospital
policy. Press
CONFIRM
once
patient status and monitoring
values have been addressed.
Press
RESTART
key per hospital
policy.
Alarm
Meaning
Response
Definitions
Alert
A visual message to help reduce programming errors by indicating a Limit
(“Soft” or “Hard”) has been exceeded. A response is required before
programming can continue.
Clinical Advisory
A visual message when a designated drug is selected, to remind clinician of
specific hospital/facility standards of practice when programming an
IV
medication. A specific Clinical Advisory and/or message can be associated
with a selected drug within any of the patient care profiles. Clinical Advisories
will not be displayed in Anesthesia mode.
Содержание Alaris 8000 Series
Страница 8: ......
Страница 52: ......
Страница 54: ......
Страница 176: ......
Страница 242: ......
Страница 284: ......
Страница 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Страница 318: ......
Страница 338: ......
Страница 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Страница 340: ......
Страница 350: ......
Страница 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Страница 352: ......
Страница 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Страница 362: ......
Страница 377: ......
Страница 378: ......