7.2 Settings for Printing and Transfer (Page: 2/3)
– 38 –
AUTO PRINT
Whether to automatically have the results of measuring both eyes
printed out in FULL AUTO mode
“
ON
”:
Results will be printed out automatically.
“
OFF
”:
switch must be pressed to print out the results.
When “
” and “
TRANS
” are set to “
OFF
”,
“
AUTO PRINT
” will also be set to “
OFF
”.
DEFAULT
Choosing whether to automatically return to AUTO mode or
FULL AUTO mode in the following cases:
- After power is turned ON
- After pressing
switch
- After deleting data by pressing
CLEAR (DEMO)
switch
- After recovering from power-saving mode
“
AUTO
”:
The instrument returns to AUTO mode.
“
FULL AUTO
”: The instrument returns to FULL AUTO mode.
(
) LEVEL
Choosing whether to display level meter on the measurement mode
screen, display mode screen and printout.
Level meter is a mark showing the waveform level of the signal
obtained by detecting light reflected from the cornea when air is
blown onto the eye.
“
ON
”:
Level meter will be displayed.
“
OFF
”:
Level meter will not be displayed.
CHARACTER
The type of characters displayed for entering message on page 3/3
can be selected. See the illustration of the monitor in section 7.3.
“
JIS
”:
Japanese characters will be displayed.
“
LATIN-1
”: Characters such as with umlaut will be displayed.
Press AUTO/MANU. switch to proceed to page 3/3.
Press PUFF TIMES switch to return to page 1/3.
Press CLEAR (DEMO) switch to return to measurement mode.
Содержание TX-F
Страница 77: ......