5.2 FULL AUTO/AUTO Measurement
– 25 –
Press
R/L
switch to measure the other eye. Alignment will be started automatically.
Or, if the other eye already has been measured, press
switch to printout the
result.
NOTE: Value may become high when the eye is covered by eyelashes or eyelid.
♦
If warning image appears
If “
WARNING IMG
” is set to
“
ON
” in SET mode, eye front
image taken at moment when the
error occurred will be displayed
on the monitor.
If the eyelid or eyelashes are
below the eyelid warning line,
take remedial action and then
press
START
switch again to
perform measurement.
Move measurement head toward the other eye
FULL AUTO mode
After the required number of data is obtained, measurement head automatically
moves toward the other eye.
AUTO mode
Press
R/L
switch in order to move the measurement head toward the other eye.
Alignment and measurement of the other eye are performed
FULL AUTO mode
Alignment and measurement are
performed automatically in the
same ways as measuring the
right eye is done when the pupil
of the left eye has been detected.
AUTO mode
Display the other eye some-
where on the monitor and press
START
switch.
Measure as many times as
required.
F U L L AU TO ( 3 )
N o . 0 0 0 0 0 1
mmHg
STD
[ , , ]
L
[ , , ]
3 0
R ERROR
Eyelid warning line
7
8
F U L L AU TO ( 3 )
N o . 0 0 0 0 0 1
mmHg
STD
[ , , ]
[ , , ]
3 0
R
11
12
12
11.7
12
14
15
13.7
L
Содержание TX-F
Страница 77: ......