CANNON Instrument Company | Communication Options
33
Table 3: Commands
Command
Type ($$)
Possible
Argument
(&..&)
Action Description
AD
None
Calibrate the 16 bit A to D converter. When complete, the
response is
AD_NO
if the calibration failed or
AD_YES
if calibration
is successful. If the command is not recognizable,
ERROR
is
displayed.
For example, if this command is sent to a CT-2000 with an address
of 5:
/5CAD
; then the response may be:
-5CADYES
RT
None
Run a complete self-test like the one run at power up. A response
immediately indicating receipt of command
RT_YES
, or the
command is unrecognizable and
ERROR
is displayed.
For example, if this command is sent to a CT-2000 with an address
of 6:
/6CRT
; then the response may be:
-6CRTYES
DC
None
Display all temperatures in °C on LCD and in all reports. This
command places the bath in Celsius mode for all temperature
references. An immediate response of command
DCYES
is
returned, or if command is unrecognizable then
ERROR
is
displayed.
For example, if this command is sent to a CT-2000 with an address
of 1:
/1CDC
; the response may be:
-1CDCYES
DF
None
Display all temperatures in °F on LCD and in all reports. This
command places the bath in Fahrenheit mode for all temperature
references. An immediate response of command
DCYES
is
returned, or if command is unrecognizable then
ERROR
is
displayed.
For example, if this command is sent to a CT-2000 with an address
of 2:
/2CDF;
then the response may be:
-2CDFYES
SR
&&&
Set the reporting interval to the value of half-seconds as defined by
the three character argument &&& in hexadecimal. Only
arguments from 000 to FFF are accepted where 000 turns all
reports off and any other numbers permit reports to be issued on
any half-second interval from 0.5 seconds up to 2,047.5 seconds,
but only if the CT-2000 is in point-to-point mode of operation. If