10
CT-2000 Instruction Manual
3.
Disconnect all external cabling including probes, heaters, and float switch from the upper rear
panel.
4.
Remove the front top cover and rear top cover. Carefully remove the rear top cover as the
temperature control probes and heating elements are attached to it, as shown in Figure 7.
Figure 7: Remove Top and Rear Top Covers
5.
Remove the glass jar from its box and packing, using caution as it is heavy.
Caution:
The glass jar is heavy. Use caution when lifting/removing jar.
6.
Ensure the rubber support ring is properly seated around the bottom of the jar opening in the
bath unit, as shown in Figure 8.
Figure 8: Checking Support Ring is Seated Properly
7.
Place the rubber gasket around the top rim of the jar. The rubber may have to be trimmed
slightly to allow the ends of the rubber gasket to meet, with
no
gap, when placed around the
rim. Refer to Figure 9.