CANNON Instrument Company | Unpack & Assemble
15
Insert Viscometer Tubes/Thermometers
The top cover of the CT-2000 contains seven apertures, 51 mm (2”) in diameter, for the insertion of
viscometer tube holders. Two additional holes are provided for insertion of thermometers. Refer to
Figure 15.
Figure 15: CT-2000 Top View shown without Hole Covers
If necessary, remove the viscometer tube cover(s) from the top of the bath and carefully place the
viscometer tube(s), with the proper holder attached, into the bath through the aperture(s) in the
top cover. The viscometer tube should be inserted to a depth which ensures that the liquid under
test and/ or any timing marks are a minimum of 6 mm (¼") below the top level of the liquid.
Thermometer Immersion
Proper thermometer immersion is critical for viscosity measurements. Even a calibrated
thermometer will read incorrectly if it is not properly immersed in the bath. Total immersion
kinematic viscosity thermometers should be used with the bulb and only the mercury column (if
applicable) beneath the surface of the liquid, but with the emergent stem above the surface at
ambient temperatures.
Note:
Different thermometers have different immersion requirements. Refer to
the information included with your particular thermometer for specific
requirements.
Filling the Bath
After CT-2000 unit assembly is complete, fill the bath.