CT-2000 Instruction Manual
Figure 25: Clear Bins Warning Message
To erase user calibration data for all integer bins, press the
ENTER
key. To exit the
Clear Bins
screen
without altering user calibration data, press the
MENU
key.
Caution:
Once the offset values are cleared they cannot be restored. The
instrument may require recalibration for all offsets to produce reliable data.
General Bath Offset
The
General Bath Offset
is a protected function of the CT-2000. This option is a global adjustment
which is made at the factory by CANNON. In rare instances when this setting may need to be
altered, the adjustment should be made in consultation with an authorized CANNON representative.