PVS5120-Series Portable Samplers
48
TABLE 8-7. Generic Modbus® Control and Monitoring of the Sampler
CRBasic
ModbusMaster()
Function Start Length
Coil
Name
Description
1 or 15
1
1
00001
Start
Start
1 or 15
2
1
00002
Serviced
Serviced
1 or 15
3
1
00003
Auto Deploy
Auto Deploy
Register*
3 or 16
1
1
40001 & 40002
Bottles
Number of bottles in sampler.
3 or 16
3
1
40003 & 40004 SampPerBottle
Number of samples to be put in each
bottle(s).
3 or 16
5
1
40005 & 40006
NeverFull
Determines if the sampler loops without
stopping (must be manually serviced before
the sampler is full to prevent overflow).
3 or 16
7
1
40007 & 40008
PurgeTime1
Initial purge time, in seconds, that the
sampler hose is purged before collecting
sample.
3 or 16
9
1
40009 & 40010 LevelSetTime
Time, in seconds, to evacuate sample
chamber to desired sample volume.
3 or 16
11
1
40011 & 40012 DepositTime Time, in seconds, to hold pinch valve open to
allow sample to be deposited in bottle.
3 or 16
13
1
40013 & 40014
PurgeTime2
Final purge time, in seconds, that the sampler
hose is evacuated after collecting sample.
3 or 16
15
1
40015 & 40016 UsePurgeTime
A true/false variable which indicates if
PurgeTime2 is set independently of
PurgeTime1 or is set to same value as
PurgeTime1.
3 or 16
17
1
40017 & 40018
SetTimeOut
True/false value that indicates whether the
vacuum time is set by the user or
automatically set (2 x PurgeTime1).
3 or 16
19
1
40019 & 40020
TimeOut
Vacuum timeout, in seconds, to determine
how long sampler should attempt to collect a
sample before triggering an error.
3 or 16
21
1
40021 & 40022 SamplerType
The sampler type determined by the
corresponding integer value:
0 = PakBus, 1 = Pulse, 2 = Modbus,
3 = SDI-12
3
23
1
40023 & 40024
Bottle
Indicates bottle number that the sample is
deposited in (1 to 24). Value is always 1 for
composite sampler. (Read only.)
3
25
1
40025 & 40026 SampleCount
Displays the number of samples deposited in
to this bottle. Value is always 1 for never full
option. (Read only.)
3
27
1
40027 & 40028
Response
Displays numeric code ranging from 0 to 8
that represents a different condition (TABLE
). 0 indicates no errors. (Read only.)
3
29
1
40029 & 40030
Batt
Returns the sampler’s battery voltage just
before that sample was collected. (Read
only.)
*The Registers contain 15 values encoded as IEEE 754, 4-byte floating point (CDAB)
Содержание PVS5120 Series
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 70: ...Appendix A Sample Transport Velocity A 2...
Страница 78: ...Appendix B Example Programs B 8...
Страница 86: ...Appendix C Monitoring Sampler Status via RS 485 C 8...
Страница 96: ...Appendix G Generic Modbus Control G 4...
Страница 100: ...Appendix H ISO5667 Conformity H 4...
Страница 101: ......