PVS5120-Series Portable Samplers
35
TABLE 8-3. Numeric Response Codes of the VSC100
Numeric
Response Code
Definition
0
No Error: Successful Sample
1
All samples complete
2
Low start battery (Voltage <11.0 Vdc)
3
Low run battery
(Voltage <10.5 Vdc occurs after pumping has started)
4
Sampler vacuum timed out; no sample detected
5
Sample overflow detected (only if an overflow sensor is
installed; only available for stationary samplers)
6
No sample liquid detected (only if water detection probe
is installed and configured)
7
Arm calibration error
8
Liquid Sensing/Conductivity Rod and
Volume Control Tube are shorted
For all CRBasic dataloggers, except the CR200(X) series, the
SDI12Recorder()
instruction has the following form:
SDI12Recorder(
Destination, SDIPort, SDIAddress, “SDICommand”,
Multiplier, Offset
)
Where,
•
Destination
– For the sampler, the destination location should be a
four value array.
•
SDIPort
– Datalogger control port to which the sampler is connected.
•
SDIAddress
– This address must match the sampler’s address. The
sampler’s default address of 0 should be used unless more than one
SDI-12 device is connected to the same port. The sampler’s default
address is changed using
DevConfig
.
•
SDICommand
– The only commands recognized by the sampler are
“
M!
” and “
C!
”
.
The “
C!
” command should only be used if you want
to run SDI-12 sensors concurrently. The command needs to be
enclosed in quotes.
The PVS5120 does not support extended SDI-12 commands.
Use the default values of 1.0 and 0.0 for the
Multiplier
and
Offset
, respectively.
NOTE
Содержание PVS5120 Series
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 70: ...Appendix A Sample Transport Velocity A 2...
Страница 78: ...Appendix B Example Programs B 8...
Страница 86: ...Appendix C Monitoring Sampler Status via RS 485 C 8...
Страница 96: ...Appendix G Generic Modbus Control G 4...
Страница 100: ...Appendix H ISO5667 Conformity H 4...
Страница 101: ......