PVS5120-Series Portable Samplers
22
TABLE 7-1. PVS5120CBL-L Cable Connections
Wire
Description
Connection
Green SDI-12/Pulse Out Datalogger control port (C1, C3, C5, or C7)
Blue
Pulse Start
Connect directly to pulse device such as
a flow meter to trigger the sampler
based on a pulse (5 V TTL)
Purple
Ground
Datalogger G terminal
Gr
e
y
Ground
Connects to one lead of the optional
water detection probe
Yellow
Water Sensor
Connects to one lead of the optional
water detection probe
White
RS-485 A-
MD485 RS-485 A port
Black
RS-485 B+
MD485 RS-485 B port
Red
Fault
Outputs 5 V TTL
Brown
RG
Resistive ground (
⏚
)
7.2.3.2 USB Cable (pn 17648)
This USB cable is shipped with the sampler (FIGURE
). It connects the
sampler to a PC, which is required for sampler configuration (see Section
Connecting to the VSC100/Sampler via DevConfig
(p. 9)
). Drivers for this cable
should install automatically (no disk necessary) the first time it is plugged into
a PC.
Be sure to wait enough time for the drivers to install before
opening Campbell Scientific’s
DevConfig
.
FIGURE 7-6. USB Cable (pn
#
17648)
7.2.3.3 External Battery Cable (pn
#
30819)
The external battery cable (pn
#
30819) connects an external battery to the
sampler via the
Power
port on the Control Panel (FIGURE
). This cable is
optional.
NOTE
Содержание PVS5120 Series
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 70: ...Appendix A Sample Transport Velocity A 2...
Страница 78: ...Appendix B Example Programs B 8...
Страница 86: ...Appendix C Monitoring Sampler Status via RS 485 C 8...
Страница 96: ...Appendix G Generic Modbus Control G 4...
Страница 100: ...Appendix H ISO5667 Conformity H 4...
Страница 101: ......