PVS5120-Series Portable Samplers
51
1. Once connected to the VSC100, click the
Sampler Settings
tab (FIGURE
).
a. Set the
Control Type
to
Pulse
.
b. Set the
Number of Sampler Bottles
.
c. Set the
Samples Per Bottle
. It can range from
1
to
999
. This setting
is ignored if
Never Full
is set to
On/Enable
.
d. Typically leave the
Never Full Operation
disabled.
e. If the Auto Deploy routine has been run, the
Purge Time
will already
be set. If not using Auto Deploy, enter the amount of time the
sampler hose is evacuated before and after the sample is collected.
The maximum value is 800 s.
f. If a
Water Present Sensor
is present, click to install. This sensor is
wired to the sampler control cable. See Section
(p. 24)
, for more information.
FIGURE 8-17. Pulse Control settings
2. Click the
Advanced Sampler Settings
tab (FIGURE
). In most cases,
the only thing a user will want to change is to Enable Pressure Relief
setting; however, if you are interested in why the other options should be
changed from factory defaults, see Section
(p. 13)
.