OBS500 Smart Turbidity Meter with ClearSensor
®
Technology
29
Short Cut
cannot edit programs after they are imported and edited
in
CRBasic Editor
.
A
Short Cut
tutorial is available in Section
. If you wish to
import
Short Cut
code into
CRBasic Editor
to create or add to a customized
program, follow the procedure in Appendix
(p. A-1)
.
Programming basics for CRBasic dataloggers are provided in the following
sections. Complete program examples for select CRBasic dataloggers can be
found in Appendix
(p. B-1)
. Programming basics and
programming examples for Edlog dataloggers are provided at
www.campbellsci.com/old-manuals
7.6.1 SDI-12 Programming
The
SDI12Recorder
instruction is used to read the OBS500 in SDI-12 mode.
A multiplier of 1.0 and an offset of 0.0 yield water level in psig and
temperature in degrees C.
The
SDI12Recorder
instruction has the following form:
SDI12Recorder
(Destination, Output String, Multiplier, Offset)
Refer to Appendix
, for an example of
using this CRBasic instruction.
7.6.2 RS-232 Programming
The
SerialOut()
instruction sends strings over the Tx COM port and the
SerialIn()
instruction receives strings from the Rx COM port.
Refer to Appendix
, for an example of
using these CRBasic instructions.
7.6.3 Analog Programming
The
PortSet
instruction is used to open the shutter. Either the
VoltDiff
(recommended) or
VoltSe
instruction is used to measure the analog voltage
output.
Refer to Appendix
, for an example of
using these CRBasic instructions.
7.7 Operation in High Sediment Loads and Sandy Sediments
Sites with high sediment loads and large sand grains can be problematic for the
shutter and its motor. The recommendations provided in this section should
help reduce these problems.
Typically sites with high biological growth have relatively low
sediment loads.
NOTE
NOTE
Содержание OBS500
Страница 2: ......
Страница 6: ......
Страница 10: ...Table of Contents iv ...
Страница 15: ...OBS500 Smart Turbidity Meter with ClearSensor Technology 5 FIGURE 4 4 Place and secure mounting fixture ...
Страница 58: ...Appendix B Example Programs B 6 ...
Страница 60: ...Appendix C OBS500 Copper Sleeve Kit Installation C 2 ...
Страница 63: ......