OBS500 Smart Turbidity Meter with ClearSensor
®
Technology
20
TABLE 7-7. SDI-12 and RS-232 Measurement Commands
for OS Version 2 or Higher
Commands
Process
Values Returned
aM7!
aC7!
Close Wiper
Send Data
close wiper position count
close max current count
close timeout count
close current (mA)
total open/close count
aC8!
Raw Burst Data in Volts
Open
Measure 100 Times
Close
Send Data
bs median
bs mean
bs Standard deviation
bs minimum
bs maximum
ss median
ss mean
ss standard deviation
ss minimum
ss maximum.
aM9!
Wiper
Leave Open
total open/close count
open wiper position count
open max current count
open timeout count
close wiper position count
close max current count
close timeout count
7.4.1.1 SDI-12 Addresses
The OBS500 SDI-12 address can be set to 0 to 9, A to Z, or a to z which allows
multiple sensors to be connected to a single digital I/O channel (control port) of
an SDI-12 datalogger. (Most Campbell Scientific dataloggers support SDI-12.)
The OBS500 is shipped from the factory with the address set to 0. When it is
necessary to measure more than one OBS500, it is easiest to use a different
control port for each OBS500 instead of changing the address. If additional
control ports are not available, the address will need to be changed.
The address on the OBS500 can be changed by sending the SDI-12 change
address command
aAb!
. The change address command can be issued from
most SDI-12 recorders. For example, to change the address of a sensor that has
a default address of 0 to the address of 1 the following command can be sent:
0A1!
The address may also be changed by connecting to the probe in
DevConfig
.
Once connected, in the Settings Editor tab click in the address box and enter
the new address. Press Apply to save the changes.
7.4.1.2 SDI-12 Transparent Mode
The transparent mode allows direct communication with the OBS500. This
may require waiting for programmed datalogger commands to finish before
sending responses. While in the transparent mode, datalogger programs may
not execute. Datalogger security may need to be unlocked before the
transparent mode can be activated.
Содержание OBS500
Страница 2: ......
Страница 6: ......
Страница 10: ...Table of Contents iv ...
Страница 15: ...OBS500 Smart Turbidity Meter with ClearSensor Technology 5 FIGURE 4 4 Place and secure mounting fixture ...
Страница 58: ...Appendix B Example Programs B 6 ...
Страница 60: ...Appendix C OBS500 Copper Sleeve Kit Installation C 2 ...
Страница 63: ......