OBS500 Smart Turbidity Meter with ClearSensor
®
Technology
21
The transparent mode is entered while the PC is in telecommunications with
the datalogger through a terminal emulator program. It is most easily accessed
through Campbell Scientific datalogger support software, but it is also
accessible with terminal emulator programs such as Windows Hyperterminal.
To enter the SDI-12 transparent mode, enter the terminal emulator from
LoggerNet
,
PC400
, or
PC200W
datalogger support software. A terminal
emulator screen is displayed. Click the
Open Terminal
button.
For CR800 series, CR1000, and CR3000 dataloggers, press <Enter> until the
datalogger responds with the prompt (e.g., “CR1000>” for the CR1000). Type
SDI12
at the prompt and press <Enter>. In response, the query
Enter Cx Port
1,3,5 or 7
will appear. Enter an integer value indicating the control port to
which the OBS500 is connected. A response of
Entering SDI12 Terminal
indicates that SDI-12 Transparent Mode is active. Any of the SDI-12
commands may be entered (e.g.,
aM1!
where
a
refers to the address). After
entering a command, the results may be viewed by entering
aD!
.
FIGURE 7-4. Terminal Emulator
Only one sensor of the same address can be connected when using the change
address command.
7.4.2 RS-232
RS-232 measurements of the OBS500 are typically made by a CR800, CR850,
CR1000, or CR3000 datalogger or an RTU device. The OBS500 field cable is
Содержание OBS500
Страница 2: ......
Страница 6: ......
Страница 10: ...Table of Contents iv ...
Страница 15: ...OBS500 Smart Turbidity Meter with ClearSensor Technology 5 FIGURE 4 4 Place and secure mounting fixture ...
Страница 58: ...Appendix B Example Programs B 6 ...
Страница 60: ...Appendix C OBS500 Copper Sleeve Kit Installation C 2 ...
Страница 63: ......