OBS500 Smart Turbidity Meter with ClearSensor
®
Technology
19
TABLE 7-7. SDI-12 and RS-232 Measurement Commands
for OS Version 2 or Higher
Commands
Process
Values Returned
aM!
aC!
a = address
Open Wiper
Measure
Close
Send Data
obs (tu)
ss (tu)
temperature (ºC)
wet dry (0=dry 1=wet)
aC1!
Burst Data
Open Wiper
Measure 100 Times
Close
Send Data
bs median
bs mean
bs standard deviation
bs minimum
bs maximum
ss median
ss mean
ss standard deviation
ss minimum
ss maximum.
aM2!
aC2!
Open Wiper
Measure
Close
Send Data
obs (tu)
ss (tu )
ratio (tu)
temperature (ºC)
raw obs (V)
raw ss (V)
open current (mA)
close current (mA)
wet dry (0=dry 1=wet)
aM3!
aC3!
Open Wiper
Send Data
open wiper position count
open max current count
open timeout count
open current (mA)
total open/close count
aM4!
aC4!
Measure
Send Data
obs (tu)
ss (tu)
temperature (ºC)
wet dry (0=dry 1=wet)
aC5!
Burst Data
Measure 100 Times
Send Data
bs median
bs mean
bs standard deviation
bs minimum
bs maximum
ss median
ss mean
ss standard deviation
ss minimum
ss maximum.
aM6!
aC6!
Measure
Send Data
obs (tu)
ss (tu )
ratio (tu)
temperature (ºC)
raw obs (V)
raw ss (V)
open current (mA)
close current (mA)
wet dry (0=dry 1=wet)
Содержание OBS500
Страница 2: ......
Страница 6: ......
Страница 10: ...Table of Contents iv ...
Страница 15: ...OBS500 Smart Turbidity Meter with ClearSensor Technology 5 FIGURE 4 4 Place and secure mounting fixture ...
Страница 58: ...Appendix B Example Programs B 6 ...
Страница 60: ...Appendix C OBS500 Copper Sleeve Kit Installation C 2 ...
Страница 63: ......