OBS500 Smart Turbidity Meter with ClearSensor
®
Technology
16
There are three settings that can be changed: SDI-12 address, measurement
mode, and sidescatter ratio top. Select the desired values and press the Apply
button.
The SDI-12 address is not used while in analog mode.
7.3 Datalogger/RTU Connection
The OBS500 field cable is typically used to connect to a datalogger or RTU.
The field cable is a molded-cable assembly that terminates with an MCIL wet
pluggable underwater terminator. TABLE
shows the contact numbers for
the MCIL/MCBH-8 connectors and the electrical functions and wire colors.
TABLE 7-3. OBS500 Connector Pin-Out
MCIL-8-MP/MCBH-8-FS
Contact Number
Electrical Function
Wire Color
1
Power Ground
Black
2
SDI-12/RS232 TX/Analog
SS-BS Control
White
3
Power (9.6 to 15 V)
Red
4
Analog Signal
Green
5
RS-232 RX/Shutter Open
Blue
6
NC
7
Analog Ground
Brown
8
NC
No Connection
Clear/Braid
This document provides the recommended wiring configuration for connecting
the OBS500 field cable to a Campbell Scientific datalogger. Wiring to
dataloggers or RTUs manufactured by other companies is similar.
Campbell Scientific recommends powering down the system
before wiring the OBS500. The shield wire plays an important role
in noise emissions and susceptibility as well as transient
protection.
NOTE
NOTE
Содержание OBS500
Страница 2: ......
Страница 6: ......
Страница 10: ...Table of Contents iv ...
Страница 15: ...OBS500 Smart Turbidity Meter with ClearSensor Technology 5 FIGURE 4 4 Place and secure mounting fixture ...
Страница 58: ...Appendix B Example Programs B 6 ...
Страница 60: ...Appendix C OBS500 Copper Sleeve Kit Installation C 2 ...
Страница 63: ......