OBS500 Smart Turbidity Meter with ClearSensor
®
Technology
26
FIGURE 7-7. OBS500 in 500-TU AMCO Clear
®
turbidity standard in
100-mm black polyethylene calibration cup
7.5.2 Sediment
There are three basic ways to calibrate an OBS sensor with sediment. These are
described in the following sections. However, only the procedures for dry-
sediment are explained in this manual. Typically, the sensor will record in
turbidity units and the relationship to suspended sediment is calculated in a
spread sheet or database after the data is retrieved to a computer.
7.5.2.1 Dry-Sediment Calibration
Dry-sediment calibration is a calibration performed with sediment that has
been dried, crushed, and turned to powder. This is the easiest calibration to do
because the amount of sediment can be determined accurately with an
electronic balance and the volume of water in which it is suspended can be
accurately measured with volumetric glassware. Of the three methods, dry-
sediment calibration causes the greatest physical and chemical alteration of the
sediment. Alteration of the sediment size as a result of processing can
significantly affect the calibration slope. FIGURE
reducing the grain size by a factor of two during grinding can increase OBS
sensitivity by a factor of two.
7.5.2.2 Wet-Sediment Calibration
Wet-sediment calibration is performed with sediment obtained from water
samples or from the bed of a river that has not been dried and pulverized.
Consolidation and biochemical changes during storage and processing cause
some alteration of wet sediment, and for this reason, sediment and water
samples should be stored at about 4 °C prior to use. The wet sediment is
introduced into the sediment suspender as it comes from the field. This kind of
Содержание OBS500
Страница 2: ......
Страница 6: ......
Страница 10: ...Table of Contents iv ...
Страница 15: ...OBS500 Smart Turbidity Meter with ClearSensor Technology 5 FIGURE 4 4 Place and secure mounting fixture ...
Страница 58: ...Appendix B Example Programs B 6 ...
Страница 60: ...Appendix C OBS500 Copper Sleeve Kit Installation C 2 ...
Страница 63: ......