Appendix C. EasyFlux
®
DL CR6KH20
C-4
4.
Once connected, a list of settings is shown. Navigate to
CPI Address
and
change the value to
1
. Press
Apply
and exit the software.
5.
Use an Ethernet cable (included with the module) to connect the module
CPI port to the CR6 CPI port.
C.3.2.2 Barometer
A CS106 Barometer is recommended for increased accuracy due to
calculations and unit conversions that use ambient pressure. TABLE
shows the default wiring for
EasyFlux DL CR6KH20.
TABLE C-2. Default Wiring for CS106 Barometer
Sensor
Quantity
Wire Description
Color
CR6 Terminal
CS106 Barometer
0 or 1
Signal
Blue
U7
Signal Reference
Yellow
⏚
Shield
Clear
⏚
12V
Red
12V
Power Ground
Black
G
Trigger (not used)
Green
G
C.3.2.3 Fine Wire Thermocouple
A fine wire thermocouple is recommended for a more accurate and direct
measurement of sensible heat flux. If no fine wire thermocouple is used, an
estimate of sensible heat flux is still given; it is derived using the covariance of
sonic temperature and vertical wind and applying the SND correction. The
EasyFlux DL CR6KH20
can support from zero to one fine-wire thermocouple.
Shown in TABLE
are the available types and default wiring for adding a
fine-wire thermocouple.
TABLE C-3. Default Wiring for Fine Wire Thermocouple
Sensor
Quantity Wire Description Color
VOLT116
Terminal
FW3 Fine Wire
Thermocouple
1/
0 or 1
Signal
Purple
Diff 15H
Signal Reference
Red
Diff15L
Shield
Clear
⏚
1/
The FW05 and FW1 may be used instead of the FW3, although they are more fragile and may require more
frequent replacement.