Appendix C. EasyFlux
®
DL CR6KH20
C-21
through TABLE
give a description of all data fields found
in each data output table and when each data field is included in the table.
Prior to coordinate rotations, the orthogonal wind components
from the sonic anemometer are denoted as U
x
, U
y
, and U
z
.
Following coordinate rotations, the common denotation of u, v,
and w is used, respectively.
Variables with _R denote that the value was computed after
coordinate rotations were done. Variables with a _F denote that
the value was calculated after frequency corrections were applied.
Similarly, _SND and _WPL refer to variables that have had the
SND correction or the WPL correction applied, respectively.
TABLE C-17. Data Fields in the Time_Series Data Output Table
Data Field
Name
Units
Description
Data Field
Included
Ux
m·s
-1
Wind speed along sonic x-axis
Always
Uy
m·s
-1
Wind speed along sonic y-axis
Always
Uz
m·s
-1
Wind speed along sonic z-axis
Always
T_SONIC
deg
C
Sonic temperature
Always
diag_sonic
none Raw sonic diagnostic value (0 indicates no diagnostic flags set)
Always
volt_KH20
mV Raw signal voltage from KH20
Always
diag_KH20
none
Diagnostic value from KH20 (a non-zero result indicates
volt_KH20 was returned as not a number, less than -10 mV, or
greater than 4500 mV).
Always
PA
kPa Ambient pressure
Always
TA_1_1_1
deg
C
Air temperature measured by the temp/RH probe
Always
RH_1_1_1
%
Relative humidity measured by the temp/RH probe
Always
FW
deg
C
Air temperature measured by fine-wire thermocouple
If FW05, FW1, or
FW3 is used
NOTE
NOTE