Appendix H. Tripod CS110 and StrikeGuard Site
H.1.2 Determination of C
site
Surface mounted upward facing CS110: SN1022 with M
parallel_plate
= 88.31.
CM10 Tripod Mounted StrikeGuard and CS110: SN1023 with M
parallel_plate
= 81.77.
Each CS110 recorded one minute averages of 1 second measurements of electric field
data for the same one hour time period from 3 AM to 4 AM on August 11, 2005. The
data from both units is plotted in Figure H-6. A best-fit line was computed. The
linear regression yields a C
site
= 0.108 for the Tripod CS110 and StrikeGuard Site.
Site Correction for CS110/StrikeGuard on CM10 Tripod August 11, 2005
1 minute average of 1 second samples
Results indicate Csite = 0.108
y = 0.1079x - 64.062
R
2
= 0.9994
-4000
-2000
0
2000
4000
6000
8000
10000
-40000
-20000
0
20000
40000
60000
80000
100000
Uncorrected Electric Field (volts/meter) for CM10 tripod mounted StrikeGuard & 2 Meter CS110
SN1023: Mparallel_plate = 81.77 Volt/meter*millivolt
E
le
c
tr
ic
fi
e
ld
(v
o
lts
/m
e
te
r) f
o
r u
p
w
a
rd
f
a
ci
n
g
C
S
1
1
0
S
N
1
0
2
2
:
Mpa
rallel_p
late
= 8
8
.31 V
o
lt/m
eter*
m
illivolt
StrikeGuard Uncorrected E_field 1_Min_Avg volts/m
Linear (StrikeGuard Uncorrected E_field 1_Min_Avg volts/m)
FIGURE H-6. Determination of C
site
H-6
Содержание CS110
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 10: ...iv This is a blank page ...
Страница 37: ...Instruction Manual 27 ...
Страница 48: ......
Страница 52: ...Appendix B CS110 Accessories This is a blank page B 2 ...
Страница 56: ...Appendix C CS110 Connector Pin outs C 4 This is a blank page ...
Страница 72: ...Appendix F Example CRBasic Programs This is a blank page F 4 ...
Страница 78: ...Appendix G CS110 2 Metre CM10 Tripod Site This is a blank page G 6 ...
Страница 85: ......