SECTION 7. MEASUREMENT PROGRAMMING EXAMPLES
7-18
TIME OUT, PARAMETER 5
The "time out", Parameter 5, specifies the
maximum length of time the instruction waits on
each repetition to receive the number of cycles
specified in Parameter 4. The time out units are
0.01 seconds. The minimum time out is the
time required to receive the specified number of
cycles at the maximum expected frequency.
Time out calculations using a recommended
9000 and 5000 cycles for temperature and
pressure at the maximum frequency are shown
below.
Time out for temperature: 6,
5.22 = (5.8
∗
10
-6
)(9000/0.01)
Time out for pressure: 16,
15.5 = (3.1
∗
10
-5
)(5000/0.01)
If the time out expires before the requested
number of cycles are read, -99999 is stored in
the input location (Parameter 6). After reading
the correct number of cycles, program
execution advances to the next instruction; the
time out may be over estimated without losing
execution speed.
CONVERTING MICROSECONDS TO
°
C AND PSI
Instruction 64 (Paroscientific Processing)
reports the temperature in
°
C and the pressure
in PSI using the measurements made by
Instruction 27 (period average).
The pressure (P) and temperature (T) in
microseconds are converted to psi and
°
C
according to the following relationships provided
by Paroscientific.
P = C (1-T
0
2
/Tau
2
)(1-D(1-T
0
2
/Tau
2
))
T = Y
1
U + Y
2
U
2
+ Y
3
U
3
where:
C
= C
1
+ C
2
U + C
3
U
2
D
= D
1
+ D
2
U
T
0
= T
1
+ T
2
U + T
3
U
2
+ T
4
U
3
+ T
5
U
4
U
= U
t
- U
0
Tau= measured period, in microseconds, of
pressure
U
t
= measured period, in microseconds, of
temperature
Values for the calibration coefficients (U
0
, Y
1
,
Y
2
, Y
3
, C
1
, C
2
, C
3
, D
1
, D
2
, T
1
, T
2
, T
3
, T
4
, T
5
)
are provided by Paroscientific.
ENTERING THE COEFFICIENTS
Coefficients are entered using CR23X
Instruction 65 (Bulk Load). A calibration sheet
from Paroscientific which accompanies the
transducer lists the 14 coefficients. Table
7.16-3 shows the correct format for entering the
coefficients. The coefficients shown are for
transducer Serial Number 30135. Your
coefficients will be different.
TABLE 7.16-3 Coefficient Entry Format for
Paroscientific "T" Series Pressure
Transducer Instruction 30
Coeff.
Value
Entry
U
0
5.860253
5.8603
Y
1
-3970.348
-3970.3
Y
2
-7114.265
-7114.3
* Y
3
102779.1
102.78
C
1
70.29398
70.294
C
2
6.610141
6.6101
C
3
-119.2867
-119.29
* D
1
0.0308837
30.884
D
2
0.0
0.0
T
1
26.33703
26.337
T
2
0.8516985
0.85170
T
3
21.80118
21.801
T
4
0.0
0.0
T
5
0.0
0.0
* Y
3
and D
1
coefficients are entered as Y
3
/1000
and D
1
*1000.
Only the first 5 digits of each 7 digit coefficient
are entered in the datalogger. The maximum
error that occurs due to using five digit
coefficients is 0.001 psi throughout a 15 psi
range at any temperature between -30 and
+60
°
C. The maximum error in temperature
calculations is 0.2
°
C over the same temperature
range.
CONNECTIONS
Figure 7.16-1 details the components required
for connecting the transducer to the CR23X.
The user supplied components are commonly
available at commercial electronic stores.
Содержание CR23X
Страница 8: ...CR23X TABLE OF CONTENTS vi This is a blank page ...
Страница 12: ...CR23X MICROLOGGER OVERVIEW OV 2 1 2 3 A 4 5 6 B 7 8 9 C 0 D FIGURE OV1 1 CR23X Micrologger ...
Страница 34: ...CR23X MICROLOGGER OVERVIEW OV 24 This is a blank page ...
Страница 50: ...SECTION 1 FUNCTIONAL MODES 1 16 This is a blank page ...
Страница 72: ...SECTION 4 EXTERNAL STORAGE PERIPHERALS 4 8 This is a blank page ...
Страница 88: ...SECTION 6 9 PIN SERIAL INPUT OUTPUT 6 10 This is a blank page ...
Страница 103: ...SECTION 7 MEASUREMENT PROGRAMMING EXAMPLES 7 15 CR23X AVW1 FIGURE 7 15 2 Well Monitoring Example ...
Страница 110: ...SECTION 7 MEASUREMENT PROGRAMMING EXAMPLES 7 22 This is a blank page ...
Страница 134: ...SECTION 8 PROCESSING AND PROGRAM CONTROL EXAMPLES 8 24 This is a blank page ...
Страница 164: ...SECTION 9 INPUT OUTPUT INSTRUCTIONS 9 30 This is a blankpage ...
Страница 188: ...SECTION 11 OUTPUT PROCESSING INSTRUCTIONS 11 8 This is a blankp age ...
Страница 221: ...SECTION 13 CR23X MEASUREMENTS 13 21 FIGURE 13 5 1 Circuits Used with Instructions 4 9 ...
Страница 229: ...14 3 1 2 3 A 4 5 6 B 7 8 9 C 0 D FIGURE 14 3 1 CR23X Battery Pack and Panel ...
Страница 239: ...SECTION 14 INSTALLATION AND MAINTENANCE 14 13 S A N Y O FIGURE 14 11 6 Removal of band clamp and battery ...
Страница 240: ...SECTION 14 INSTALLATION AND MAINTENANCE 14 14 This is a blank page ...
Страница 244: ...APPENDIX A GLOSSARY A 4 This is a blank page ...
Страница 268: ...APPENDIX B CONTROL PORT SERIAL I O INSTRUCTION 15 B 24 This is a blank page ...
Страница 276: ...APPENDIX C BINARY TELECOMMUNICATIONS C 8 This is a blank page ...
Страница 278: ...This is a blank page ...
Страница 282: ...APPENDIX F DYNAGAGE SAP FLOW P67 F 4 This is a blank page ...
Страница 299: ...APPENDIX I TD OPERATING SYSTEM ADDENDUM FOR CR510 CR10X AND CR23X MANUALS ...
Страница 300: ...This is a blank page ...
Страница 302: ...This is a blank page ...
Страница 308: ...TABLE DATA ADDENDUM AD 6 This is a blank page ...
Страница 324: ...TD ADDENDUM SECTION 1 FUNCTIONAL MODES AD 1 8 This is a blank page ...
Страница 340: ...TD ADDENDUM SECTION 8 PROCESSING AND PROGRAM CONTROL EXAMPLES AD 8 10 This is a blank page ...
Страница 342: ...This is a blank page ...
Страница 373: ...This is a blank page ...