5.5 Backup and restore
This process backs up the program, configuration, and calibration data. It is only applicable when
all the connections configurations remain unchanged. The backup and restore process requires a
computer running the
Device Configuration utility
software to be connected to the data logger
USB port.
Backup procedure:
Create a backup after the AeroX Light RLIM has been fully configured and all current transducers
have been successfully calibrated.
1. Connect the data logger USB port to your computer using the supplied USB cable.
2. Access the
Device Configuration Utility
software.
3. Type CR1000X in the device type window and select the CR1000X Series.
4. For Connection Type, click Direct.Do not check Specify PakBus Address.
5. Select Backup -> Back Up Datalogger to launch the backup function in the
Device
Configuration
utility.
6. Create the Output File Name and browse the local computer to select the directory to
place the backup file. Campbell Scientific recommends changing the default name to the
program name with the backup date (for example, CsiRLIMmsg_V1.0_1-1-22). The file name
of the program running on the data logger can be found in the Status Table. For example, if
the Status Table displays ProgName | CPU:CsiRLIMmsg_V1.0.CR1X, the CR1000X data
logger is running a data logger program named CsiRLIMmsg..
7. Click Next
8. Select the program file (.CR1X) and the calibration file (.cal)