LEFM 3xxCi User Manual
IB1406 Rev 12
Section 7
Page 52
February 2017
8.
Screw the front lid of the enclosure back on. This should be hand tight with at least 8 full turns of
the lid. Tighten the 2 mm hex set screw on the lid.
Following replacement of the CTC board, the “safe mode” readout may appear in the transmitter
display. This is an indicator that the configuration data has not been uploaded to the transmitter.
See Reprogramming the Transmitter in Section 8 for instructions on downloading the
configuration file.
7.6
MXR Replacement
Never open the transmitter when the instrument is energized. Before inspecting components,
open the power supply circuit breaker. Failure to do so may result in electrical shock and/or an
explosion if in a hazardous area.
If any component on the MXR assembly fails, the entire assembly is to be replaced. To replace the MXR,
perform the following steps, referring to Figure 7.6 for hardware locations.
Phase 1 – Disconnecting the Ribbon Cable from the CTC
Note – The ribbon cable must be removed from the CTC card to disassemble the enclosure. So the first
steps are merely to remove the ribbon cable. Follow Section 7.5, steps 1 through 3 in order to disconnect
the ribbon cable.
Phase 2 – Removing the Top Portion of the Transmitter to Expose the MXR
WARNING
CAUTION