LEFM 3xxCi User Manual
IB1406 Rev 12
Section 7
Page 45
February 2017
Maintenance
Service should be performed on the LEFM 3xxCi only by qualified personnel.
Introduction
The troubleshooting and maintenance procedures in this section may be incorporated into the customer’s
standard maintenance program. The procedures should be performed only by a trained maintenance
technician. For additional assistance from Cameron, please contact:
Cameron Ultrasonics Technology Center
1000 McClaren Woods Drive
Coraopolis, PA 15108 USA
Tel 724-273-9300
7.1
General Inspections – Preventative Maintenance Procedures
Wear an ESD protective wrist strap to avoid damaging any components.
Never open the transmitter or the meter body manifold when the instrument is energized. Before
inspecting components, open the power supply circuit breaker. Failure to do so may result in
electrical shock and/or an explosion.
The following procedure covers the inspection of the transmitters, transducers, and metering sections.
Enclosure Inspection (perform bi-annually or according to site guidelines)
Perform the following inspections on each enclosure:
1.
Verify that the transmitter enclosure and the meter body have suffered no structural damage.
Report any damage to the proper maintenance supervisor.
2.
Remove dust, dirt, and other soiling from the enclosure. Use a damp cloth to clean surfaces.
3.
Remove access covers.
a.
Inspect gaskets. Clean gaskets and mating surfaces on the enclosure with water if they
are dirty.
b.
Contact Cameron if there is any corrosion on the mating surfaces.
WARNING
CAUTION
WARNING