LEFM 3xxCi User Manual
IB1406 Rev 12
Section 6
Page 43
August 2017
Display 6 —Velocity Data
(Two screens, first for paths 1 through 4 and the second for paths 5 through 8)
Figure 6.7 Display 6 – Velocity Data
Path Alarm Conditions
The following codes (described in the LEFM Modbus manual) are used to indicate the status of each
ultrasonic path:
•
Normal - Path is operating normally
•
Reject -Path is rejecting data due to low signal-to-noise ratio, irregular statistics, or failing cross-
correlation tests
•
VOS - Path sound velocities are inconsistent with thresholds (typically, a spread of 0.15% or less
between paths is acceptable)
•
VOT - Path velocity is inconsistent at low flow rates