Other Alarm Conditions
Symptom
Possible Cause
Your Action
While in pulse
mode, the
compressor speeds
up and the pulse
mode LED turns on.
The Eclipse 5 is
unable to detect
the user’s breathing
effort. See page
18 for pulse mode
information.
1) Pulse mode may not work correctly if the user has any condition
that causes blocked/restricted nasal passages. (Illness, deviated
septum, etc.)
2) Ensure cannula is no longer than 7 feet in length.
3) Ensure that there are no kinks or blockages in cannula tubing.
4) If a humidifier bottle is connected, remove the humidifier bottle
and connect cannula directly to outlet port. Pulse mode will not
work with a humidifier bottle installed.
5) If the condition persists, contact your healthcare provider.
Power Cartridge
is draining while
plugged in to AC/
DC power supply
The Eclipse 5 is not
receiving power
from the attached
power supply.
1) Verify that the outlet is providing power. Do not connect the
power supply to a dimmer circuit or a power strip.
2) Check that cable connections on power supplies are secure to
the wall/vehicle and concentrator. If using the AC power supply,
check the cable connection on the power supply transformer
as well.
3) Ensure that the power verification lights are on.
- There will be a green light on the power supply box.
- There will be a green light on the Eclipse 5 concentrator control
panel in the shape of a power cord.
4) If power verification LEDs are not lit, remove all connections of
the power supply for 20 seconds and reconnect.
5) If the condition persists, contact your healthcare provider.
The Power
Cartridge (battery)
charge is not
lasting as long as it
should.
The power cartridge
was not fully
charged
1) Connect to AC power to recharge the battery. Verify that the
battery charges for 2-5 hours and the battery icon is full and
not flashing before use.
2) Refer to the actions for “Power Cartridge is Not Charging”
Symptom below.
The Eclipse 5 was
not operating at the
anticipated flow
rate.
1) Ensure that you are using the flow rate prescribed by your
physician and that pulse or continuous flow is being used as
required.
The power cartridge
is not preforming to
specifications.
1) Attempt to drain the battery cartridge completely. Do this by
running the Eclipse 5 on battery power until the machine shuts
off completely. Then remove the battery and allow it to cool for
approximately 30 minutes. After this time, re-insert the battery
in the Eclipse 5 and recharge it using AC power.
2) If the battery still does not last as long as it should after follow-
ing step 1, contact your healthcare provider.