CAIRE Eclipse 5 Скачать руководство пользователя страница 16

16 - ENG    PN 21182453 C | User Manual

Eclipse 5

Rechargeable Power Cartridge (Battery)

The Eclipse 5 can be powered by the rechargeable 

Power Cartridge (Battery), which is supplied with 

the device. 

Installing the Battery: Align the Battery so that it 

will slide into the empty Power Compartment. Push 

the Battery into the empty Power Compartment 

until it locks (clicks) into position and is flush with 

the back of the device. When the Battery is properly 

installed and the Eclipse 5 is ON, the Battery Status 

Gauge will appear on the Control Panel. (If you DO 

NOT have a Battery installed, the Power Cartridge 

Status Gauge will not illuminate). The unit will still 

operate on external power.

To remove the Battery, pull down on the black re-

lease lever. While holding DOWN the release lever, 

pull the Battery away from the Eclipse 5.

Battery Operating Times: The duration-of-use of 

the Battery is displayed at the top of the Control 

Panel. A variety of factors, such as flow setting, 

Pulse or Continuous Flow Mode, and breath rate 

will impact the operating time. The table Typical 

New Power Cartridge Operating Times provides 

operating time estimates for the Eclipse 5 using a 

new, fully charged Battery based on flow settings 

and operating conditions as indicated. 

Rechargeable Power 

Cartridge (Battery)

Item#7082-SEQ

WARNING: FOR OXYGEN CONCEN-
TRATORS EQUIPPED WITH BATTERIES: 
STORE IN A COOL AND DRY LOCATION 
TO HELP ENSURE THE LONGEVITY OF 
YOUR BATTERY.  STORING YOUR OXY-
GEN CONCENTRATOR FOR EXTENDED 
PERIODS OF TIME AT HIGH TEMPERA-
TURES OR WITH A FULLY CHARGED OR 
COMPLETELY DISCHARGED BATTERY 
CAN DEGRADE ITS OVERALL BATTERY 
LIFE. DO NOT ATTEMPT TO OPEN THE 
BATTERY; THERE ARE NO SERVICEABLE 
PARTS INSIDE THE BATTERY. KEEP 
BATTERIES AWAY FROM CHILDREN.

Initial Battery Charging

The new Battery supplied with your Eclipse 5 is not 

fully charged when it is shipped from the factory. 

Before using your Eclipse 5 for the first time, you 

must first fully charge the Battery. 

With the AC Power Supply plugged in, and the 

Battery correctly installed in the Power Compart-

ment of the device, allow the Battery to completely 

charge. The Battery is fully charged when the Pow-

er Cartridge Status Gauge, on the Control Panel, is 

no longer cascading.

The initial battery charge can occur with the 

Eclipse either on or off and connected to AC or 

DC power.  The Eclipse can be used while either 

discharging or recharging the Battery. It may take 

between 2 and 5 hours, dependent upon the flow 

setting, to achieve 80% capacity from a fully 

discharged Battery. The battery will also recharge 

when the Eclipse is plugged into AC or DC power 

and turned off.

CAUTION: When the automobile in 
which you are using the Oxygen Con-
centrator unit is turned off, discon-
nect and remove the device from the 
automobile. Do not store the Oxygen 
Concentrator in a very hot or cold 
automobile or in other similar, high-or 
low-temperature environments. DO 
NOT leave the Oxygen Concentrator 
or the Power Supply plugged into 
the vehicle if the ignition is in the 
OFF position. Doing so may drain the 
vehicle’s battery.

CAUTION: Only use the Manufacturer’s provided 
batteries. For proper battery disposal, contact 
your Equipment Provider or your local govern-
ment agency for disposal requirements.

Содержание Eclipse 5

Страница 1: ...Eclipse 5 with autoSAT User Manual English...

Страница 2: ...ial Number Reg 2498 This way up Reg 0623 Fragile handle with care Reg 0621 Keep away from rain keep dry Reg 0626 Stacking limit by number Reg 2403 ISO 7010 Graphical symbols Safety colors and safety s...

Страница 3: ...sieve This material separates the oxygen from the nitrogen and allows only the oxygen to pass through The result is a flow of high concentration oxygen delivered to the user The Eclipse 5 combines ad...

Страница 4: ...al Become familiar with the key features of the Eclipse 5 and the User Control Panel Eclipse 5 AC Power Supply DC Power Supply Universal Cart Power Cartridge battery FRONT Control Panel Handle Oxygen...

Страница 5: ...nd Indicator The button toggles between Contin uous Flow and Pulse Dose Mode The Pulse Dose Mode activates autoSAT Technology and allows a significant increase in the operating time while powered by t...

Страница 6: ...back and forth between continuous flow and pulse dose modes In Continuous Flow Mode the oxygen is provided at a constant flow rate between 0 5 and 3 0 LPM In Pulse Dose Mode ox ygen is supplied in a...

Страница 7: ...ION ABOUT THE DISCOMFORT AND OR THE MEDICAL URGENCY TO THE RESPONSIBLE CARE GIVER TO AVOID HARM PATIENTS WITH HEARING AND OR SIGHT IMPAIRMENT S MAY NEED ASSISTANCE WITH MONITORING ALARMS Contraindicat...

Страница 8: ...ENTRATOR THE BATTERY MUST ALSO BE REMOVED FROM THE DEVICE AND PACKAGED PROPERLY WARNING DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTAND ING THIS MANUAL IF YOU ARE UNABLE TO UNDERSTA...

Страница 9: ...ING THE US DEPARTMENT OF TRANS PORTATION DOT AND UNITED NATIONS UN REGULATIONS REQUIRE THE REMOVAL OF THE BATTERY FROM THE DEVICE FOR ALL INTERNA TIONAL AIRLINE TRAVEL WHEN THE OXYGEN CONCENTRATOR IS...

Страница 10: ...IF UNABLE TO LEAVE THE ROOM YOU MUST WAIT 10 MINUTES AFTER YOU HAVE TURNED THE OXYGEN CONCENTRATOR OFF BEFORE SMOKING WARNING NO SMOKING OXYGEN IN USE SIGNS MUST BE PROMINENTLY DISPLAYED IN THE HOME O...

Страница 11: ...llow these instructions can result in the power supply not supplying power to the Oxygen Concentrator CAUTION When the automobile in which you are using the Oxygen Concentrator unit is turned off disc...

Страница 12: ...t Nominal Battery Recharge Time 1 8 to 5 0 hrs recharge time to achieve 80 capacity dependent upon the flow Alarms Alerts Loss of power Low Battery Low Therapeutic O2 Output O2 flow outside normal lim...

Страница 13: ...et filter at the top rear of the cabi net and allow exhaust air to freely leave the exhaust vent at the bottom left of the device Place the device such that the alarms can be heard Position the oxygen...

Страница 14: ...326 AC POWER SUPPLY WITH TWO PINS The Eclipse 5 includes a universal AC Power Supply for use at home or wherever standard AC power is available To connect to an AC power source be sure the AC Power Su...

Страница 15: ...le from the vehicle s DC power outlet WARNING DO NOT ALLOW EITHER THE AIR INTAKE OR THE AIR OUTLET VENTS TO BECOME BLOCKED THIS CAN CAUSE THE OXYGEN CONCEN TRATOR TO OVERHEAT AND IMPAIR PERFORMANCE WA...

Страница 16: ...RES OR WITH A FULLY CHARGED OR COMPLETELY DISCHARGED BATTERY CAN DEGRADE ITS OVERALL BATTERY LIFE DO NOT ATTEMPT TO OPEN THE BATTERY THERE ARE NO SERVICEABLE PARTS INSIDE THE BATTERY KEEP BATTERIES AW...

Страница 17: ...Note Operating the Oxygen Concentrator outside of its normal operating temperature range can affect performance and decrease battery run time and or increase battery charge time Refer to the Specific...

Страница 18: ...ot detected The device will continually check for an inspiratory effort every 15 seconds Once an inspiration is detected the device reverts to measured bolus delivery To exit Pulse Dose Mode and engag...

Страница 19: ...7 feet 2 1 m of tubing PN 5408 SEQ For any additional recommended accessories please see the Accessories Catalog PN ML LOX0010 available on www CAIREmedical com A firebreak is required for use with an...

Страница 20: ...he yellow light is on Battery symbol is flashing on LCD The Eclipse 5 power cartridge voltage may be low 1 Plug Eclipse 5 into AC power supply and recharge or replace power cartridge with fully charge...

Страница 21: ...battery CHARGER ON OR OFF ON 1 alarm beep repeat every 20 seconds 001 Ambient pressure fail AMB PRESS ON OR OFF ON 1 alarm beep repeat every 20 seconds 004 O2 85 After warm up O2 LOW OFF ON 1 alarm be...

Страница 22: ...sformer as well 3 Ensure that the power verification lights are on There will be a green light on the power supply box There will be a green light on the Eclipse 5 concentrator control panel in the sh...

Страница 23: ...be a green light on the Eclipse 5 concentrator control panel in the shape of a power cord 4 If power verification LEDs are not lit remove all connections of the power supply for 20 seconds and reconne...

Страница 24: ...r in warm tap water using a mild soap detergent solution 3 Rinse the filter thoroughly with clear tap water and squeeze out the excess water 4 Allow the filter to air dry 5 Reinsert the filter in the...

Страница 25: ...BASED PROD UCTS CONCENTRATED CHLORINE BASED PROD UCTS ETHYLENE CHLORIDE AND OIL BASED PRODUCTS PINE SOL LESTOIL THESE ARE NOT TO BE USED TO CLEAN THE PLASTIC HOUS ING ON OXYGEN CONCENTRATOR AS THEY CA...

Страница 26: ...ssions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B The device is suitable for use in all establishments including domestic es...

Страница 27: ...mmended that the Eclipse 5 is powered from an uninterruptible power supply UPS Power frequency magnet ic field IEC 61000 4 8 30A m Power frequency magnetic fields should be at levels characteristic of...

Страница 28: ...28 ENG PN 21182453 C User Manual Eclipse 5 NOTES...

Страница 29: ...PN 2118245 C User Manual 29 ENG Eclipse 5 NOTES...

Страница 30: ...30 ENG PN 21182453 C User Manual Eclipse 5 NOTES...

Страница 31: ...2 3 3 4 5 5 6 7 8 Table is based on 20 Breaths Per Minute for Pulse Dose Mode only 7 1 2 2 3 3 4 5 5 6 6 8 1 2 3 3 4 5 6 6 7 8 9 1 2 3 4 5 6 7 8 8 9 Table is based on 15 Breaths Per Minute for Pulse...

Страница 32: ...2018 Chart Industries Chart Industries reserves the right to discontinue its products or change the prices materials equipment quality descriptions specifications and or processes to its products at a...

Отзывы: