14 - ENG    PN 21182453 C | User Manual

Eclipse 5

Powering on your Eclipse 5 

The device is capable of being 

operated directly from three (3) 

different power sources:

• AC Power Supply with 

NEMA Power Cord 

• DC Power Supply 

• Rechargeable Power  

Cartridge (Battery)

General 
Charging 
Information

The Eclipse 

5 charges the 

battery while  

using AC power and when the battery temperature 

is within safe charging temperature. When external 

power is disconnected, the device will automat-

ically switch over to the battery if the battery is 

installed and charged. When external power is 

restored by plugging into an AC outlet, the device 

will accept power from the external power source 

and replenish the battery.  It is not mandatory 

for the battery to be installed for the Eclipse 5 to 

operate.

Proper connection of the AC or DC Power Supply 

requires aligning the power plug with the recessed 

external power receptacle of the device. The 

recessed external power receptacle is located on 

the right side of the device when facing the Control 

Panel. Push the power plug into the recessed 

receptacle.

WARNING: THE ECLIPSE BATTERY 
CARTRIDGE AND DC POWER SUPPLY 
MUST BE PERIODICALLY CHECKED. 
WHEN AC POWER IS NOT AVAILABLE 
THESE ALTERNATE POWER SOURCES 
MAY BE USED.

AC Two-Pin Power Supply 

Item# 20852326

AC POWER SUPPLY WITH TWO PINS

The Eclipse 5 includes a universal AC Power 

Supply for use at home or wherever standard AC 

power is available. To connect to an AC power 

source, be sure the AC Power Supply cord is 

securely inserted into the recessed power receptacle 

on the side of the device and that the cord from the 

AC Power Supply is plugged into an AC outlet. 

When the Eclipse 5 is plugged in properly, a green 

indicator on the AC Power Supply will light and the 

External Power Indicator, 

 will appear on the 

user control panel.

Note: Use only electrical voltage specified on the 
specification label affixed to the device.

CAUTION: Use only accessories rec-
ommended by the manufacturer. Use 
of any other may be hazardous, cause 
serious damage to your oxygen con-
centrator and will void the warranty.

CAUTION: Do not use extension cords with 
this unit or connect too many plugs into the 
same electrical outlet. The use of extension 
cords could adversely affect the performance 
of the device. Too many plugs into one outlet 
can result in an overload to the electrical panel 
causing the breaker/fuse to activate or fire if 
the breaker or fuse fails to operate.
CAUTION: Use of cables and adapters other 
than those specified, with the exception of 
cables and adapters sold by the manufacturer 
of the medical electrical equipment as replace-
ment parts for internal components, may result 
in increased emissions of decreased immunity 
of the Oxygen Concentrator.

Step 1

Step 2

Содержание Eclipse 5

Страница 1: ...Eclipse 5 with autoSAT User Manual English...

Страница 2: ...ial Number Reg 2498 This way up Reg 0623 Fragile handle with care Reg 0621 Keep away from rain keep dry Reg 0626 Stacking limit by number Reg 2403 ISO 7010 Graphical symbols Safety colors and safety s...

Страница 3: ...sieve This material separates the oxygen from the nitrogen and allows only the oxygen to pass through The result is a flow of high concentration oxygen delivered to the user The Eclipse 5 combines ad...

Страница 4: ...al Become familiar with the key features of the Eclipse 5 and the User Control Panel Eclipse 5 AC Power Supply DC Power Supply Universal Cart Power Cartridge battery FRONT Control Panel Handle Oxygen...

Страница 5: ...nd Indicator The button toggles between Contin uous Flow and Pulse Dose Mode The Pulse Dose Mode activates autoSAT Technology and allows a significant increase in the operating time while powered by t...

Страница 6: ...back and forth between continuous flow and pulse dose modes In Continuous Flow Mode the oxygen is provided at a constant flow rate between 0 5 and 3 0 LPM In Pulse Dose Mode ox ygen is supplied in a...

Страница 7: ...ION ABOUT THE DISCOMFORT AND OR THE MEDICAL URGENCY TO THE RESPONSIBLE CARE GIVER TO AVOID HARM PATIENTS WITH HEARING AND OR SIGHT IMPAIRMENT S MAY NEED ASSISTANCE WITH MONITORING ALARMS Contraindicat...

Страница 8: ...ENTRATOR THE BATTERY MUST ALSO BE REMOVED FROM THE DEVICE AND PACKAGED PROPERLY WARNING DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTAND ING THIS MANUAL IF YOU ARE UNABLE TO UNDERSTA...

Страница 9: ...ING THE US DEPARTMENT OF TRANS PORTATION DOT AND UNITED NATIONS UN REGULATIONS REQUIRE THE REMOVAL OF THE BATTERY FROM THE DEVICE FOR ALL INTERNA TIONAL AIRLINE TRAVEL WHEN THE OXYGEN CONCENTRATOR IS...

Страница 10: ...IF UNABLE TO LEAVE THE ROOM YOU MUST WAIT 10 MINUTES AFTER YOU HAVE TURNED THE OXYGEN CONCENTRATOR OFF BEFORE SMOKING WARNING NO SMOKING OXYGEN IN USE SIGNS MUST BE PROMINENTLY DISPLAYED IN THE HOME O...

Страница 11: ...llow these instructions can result in the power supply not supplying power to the Oxygen Concentrator CAUTION When the automobile in which you are using the Oxygen Concentrator unit is turned off disc...

Страница 12: ...t Nominal Battery Recharge Time 1 8 to 5 0 hrs recharge time to achieve 80 capacity dependent upon the flow Alarms Alerts Loss of power Low Battery Low Therapeutic O2 Output O2 flow outside normal lim...

Страница 13: ...et filter at the top rear of the cabi net and allow exhaust air to freely leave the exhaust vent at the bottom left of the device Place the device such that the alarms can be heard Position the oxygen...

Страница 14: ...326 AC POWER SUPPLY WITH TWO PINS The Eclipse 5 includes a universal AC Power Supply for use at home or wherever standard AC power is available To connect to an AC power source be sure the AC Power Su...

Страница 15: ...le from the vehicle s DC power outlet WARNING DO NOT ALLOW EITHER THE AIR INTAKE OR THE AIR OUTLET VENTS TO BECOME BLOCKED THIS CAN CAUSE THE OXYGEN CONCEN TRATOR TO OVERHEAT AND IMPAIR PERFORMANCE WA...

Страница 16: ...RES OR WITH A FULLY CHARGED OR COMPLETELY DISCHARGED BATTERY CAN DEGRADE ITS OVERALL BATTERY LIFE DO NOT ATTEMPT TO OPEN THE BATTERY THERE ARE NO SERVICEABLE PARTS INSIDE THE BATTERY KEEP BATTERIES AW...

Страница 17: ...Note Operating the Oxygen Concentrator outside of its normal operating temperature range can affect performance and decrease battery run time and or increase battery charge time Refer to the Specific...

Страница 18: ...ot detected The device will continually check for an inspiratory effort every 15 seconds Once an inspiration is detected the device reverts to measured bolus delivery To exit Pulse Dose Mode and engag...

Страница 19: ...7 feet 2 1 m of tubing PN 5408 SEQ For any additional recommended accessories please see the Accessories Catalog PN ML LOX0010 available on www CAIREmedical com A firebreak is required for use with an...

Страница 20: ...he yellow light is on Battery symbol is flashing on LCD The Eclipse 5 power cartridge voltage may be low 1 Plug Eclipse 5 into AC power supply and recharge or replace power cartridge with fully charge...

Страница 21: ...battery CHARGER ON OR OFF ON 1 alarm beep repeat every 20 seconds 001 Ambient pressure fail AMB PRESS ON OR OFF ON 1 alarm beep repeat every 20 seconds 004 O2 85 After warm up O2 LOW OFF ON 1 alarm be...

Страница 22: ...sformer as well 3 Ensure that the power verification lights are on There will be a green light on the power supply box There will be a green light on the Eclipse 5 concentrator control panel in the sh...

Страница 23: ...be a green light on the Eclipse 5 concentrator control panel in the shape of a power cord 4 If power verification LEDs are not lit remove all connections of the power supply for 20 seconds and reconne...

Страница 24: ...r in warm tap water using a mild soap detergent solution 3 Rinse the filter thoroughly with clear tap water and squeeze out the excess water 4 Allow the filter to air dry 5 Reinsert the filter in the...

Страница 25: ...BASED PROD UCTS CONCENTRATED CHLORINE BASED PROD UCTS ETHYLENE CHLORIDE AND OIL BASED PRODUCTS PINE SOL LESTOIL THESE ARE NOT TO BE USED TO CLEAN THE PLASTIC HOUS ING ON OXYGEN CONCENTRATOR AS THEY CA...

Страница 26: ...ssions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B The device is suitable for use in all establishments including domestic es...

Страница 27: ...mmended that the Eclipse 5 is powered from an uninterruptible power supply UPS Power frequency magnet ic field IEC 61000 4 8 30A m Power frequency magnetic fields should be at levels characteristic of...

Страница 28: ...28 ENG PN 21182453 C User Manual Eclipse 5 NOTES...

Страница 29: ...PN 2118245 C User Manual 29 ENG Eclipse 5 NOTES...

Страница 30: ...30 ENG PN 21182453 C User Manual Eclipse 5 NOTES...

Страница 31: ...2 3 3 4 5 5 6 7 8 Table is based on 20 Breaths Per Minute for Pulse Dose Mode only 7 1 2 2 3 3 4 5 5 6 6 8 1 2 3 3 4 5 6 6 7 8 9 1 2 3 4 5 6 7 8 8 9 Table is based on 15 Breaths Per Minute for Pulse...

Страница 32: ...2018 Chart Industries Chart Industries reserves the right to discontinue its products or change the prices materials equipment quality descriptions specifications and or processes to its products at a...

Отзывы: