PN 2118245 C | User Manual    17 - ENG

Eclipse 5

  * Batteries are recyclable. *

In the event of an AC or DC power interruption, 

the Eclipse 5 will beep and automatically switch to 

Battery operation, as long as the Battery is present 

and charged. When AC power is restored, the 

Eclipse 5 will beep and the Battery will auto-

matically start recharging, if not fully charged or 

too warm. If the Battery is not present, or fully 

discharged, during an AC power interruption, the 

Eclipse 5 will shut down and alarm. The Battery is 

fully discharged when the Power Cartridge Status 

Gauge on the control panel is no longer illumi-

nated. The device will display a yellow light and 

sound a beep for 120 seconds. The yellow light will 

remain on and the beep will repeat after 20 seconds 

until the device completely shuts down.

Note: Press and hold the Eclipse 5 ON/OFF button 
for a minimum of 5 seconds, to silence the Loss 
of Power alarm. Connecting the Eclipse 5 to an 
external AC power supply for 2 minutes will also 
eliminate this alarm.

Typical New Power Cartridge Operating Times

Continuous 
Flow

Battery 
Duration

Pulse 
Setting

Battery 
Duration 
(12 BPM)

0.5 LPM

4.4 hours

-

-

1.0 LPM

3.7 hours

16mL–1.0

5.4 hours

2.0 LPM

2.0 hours

32mL–2.0

5.1 hours

3.0 LPM

1.3 hours

48mL–3.0

4.9 hours

64mL–4.0

4.0 hours

80mL–5.0

3.7 hours

96mL–6.0

3.5 hours

128mL–7

2.5 hours

160mL–8

2.0 hours

192mL–9

1.7 hours

Note: Operating the Oxygen Concentrator outside 
of its normal operating temperature range can 
affect performance and decrease battery run time 
and/or increase battery charge time. (Refer to the 
Specifications section in this manual.) You will 
achieve longer operating time on the Battery if you 
operate your Eclipse 5 in the Pulse Mode (Physician 
approved mode).

Typical Battery Recharge Time

The typical time to recharge your Battery to 

achieve 80% capacity from a fully discharged 

Battery is between 2 and 5 hours, depending upon 

the device flow setting. When the battery is fully 

charged, the battery gauge will cease cascading. 

The battery can be charged in the unit when the 

Eclipse 5 is turned off and plugged in.

If the Battery becomes too warm during discharg-

ing, recharging will not begin until the Battery suf-

ficiently cools. Removing the Battery and allowing 

it to cool may expedite this cooling process.

When operating the Eclipse 5 on DC power, the 

battery will recharge while the unit is running at 

2.0LPM or less if sufficient power is available. 

Typical Power Cartridge Operating Times at 80% 
Charge

Continuous 
Flow

Battery 
Duration 
(12 BPM)

Pulse 
Setting

Battery 
Duration 
(12 BPM)

0.5 LPM

4.0 hours

16mL–1.0

3.0 hours

1.0 LPM

3.5 hours

24mL–1.5

3.0 hours

1.5 LPM

2.5 hours

32mL–2.0

3.0 hours

2.0 LPM

2.0 hours

40mL–2.5

3.0 hours

2.5 LPM

1.5 hours

48mL–3.0

3.0 hours

3.0 LPM

1.3 hours

56mL–3.5

3.0 hours

64mL–4.0

3.0 hours

72mL–4.5

3.0 hours

80mL–5.0

2.5 hours

88mL–5.5

2.5 hours

96mL–6.0

2.5 hours

128mL–7.0 2.0 hours
160mL–8.0 1.5 hours
192mL–9.0 1.5 hours

Содержание Eclipse 5

Страница 1: ...Eclipse 5 with autoSAT User Manual English...

Страница 2: ...ial Number Reg 2498 This way up Reg 0623 Fragile handle with care Reg 0621 Keep away from rain keep dry Reg 0626 Stacking limit by number Reg 2403 ISO 7010 Graphical symbols Safety colors and safety s...

Страница 3: ...sieve This material separates the oxygen from the nitrogen and allows only the oxygen to pass through The result is a flow of high concentration oxygen delivered to the user The Eclipse 5 combines ad...

Страница 4: ...al Become familiar with the key features of the Eclipse 5 and the User Control Panel Eclipse 5 AC Power Supply DC Power Supply Universal Cart Power Cartridge battery FRONT Control Panel Handle Oxygen...

Страница 5: ...nd Indicator The button toggles between Contin uous Flow and Pulse Dose Mode The Pulse Dose Mode activates autoSAT Technology and allows a significant increase in the operating time while powered by t...

Страница 6: ...back and forth between continuous flow and pulse dose modes In Continuous Flow Mode the oxygen is provided at a constant flow rate between 0 5 and 3 0 LPM In Pulse Dose Mode ox ygen is supplied in a...

Страница 7: ...ION ABOUT THE DISCOMFORT AND OR THE MEDICAL URGENCY TO THE RESPONSIBLE CARE GIVER TO AVOID HARM PATIENTS WITH HEARING AND OR SIGHT IMPAIRMENT S MAY NEED ASSISTANCE WITH MONITORING ALARMS Contraindicat...

Страница 8: ...ENTRATOR THE BATTERY MUST ALSO BE REMOVED FROM THE DEVICE AND PACKAGED PROPERLY WARNING DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTAND ING THIS MANUAL IF YOU ARE UNABLE TO UNDERSTA...

Страница 9: ...ING THE US DEPARTMENT OF TRANS PORTATION DOT AND UNITED NATIONS UN REGULATIONS REQUIRE THE REMOVAL OF THE BATTERY FROM THE DEVICE FOR ALL INTERNA TIONAL AIRLINE TRAVEL WHEN THE OXYGEN CONCENTRATOR IS...

Страница 10: ...IF UNABLE TO LEAVE THE ROOM YOU MUST WAIT 10 MINUTES AFTER YOU HAVE TURNED THE OXYGEN CONCENTRATOR OFF BEFORE SMOKING WARNING NO SMOKING OXYGEN IN USE SIGNS MUST BE PROMINENTLY DISPLAYED IN THE HOME O...

Страница 11: ...llow these instructions can result in the power supply not supplying power to the Oxygen Concentrator CAUTION When the automobile in which you are using the Oxygen Concentrator unit is turned off disc...

Страница 12: ...t Nominal Battery Recharge Time 1 8 to 5 0 hrs recharge time to achieve 80 capacity dependent upon the flow Alarms Alerts Loss of power Low Battery Low Therapeutic O2 Output O2 flow outside normal lim...

Страница 13: ...et filter at the top rear of the cabi net and allow exhaust air to freely leave the exhaust vent at the bottom left of the device Place the device such that the alarms can be heard Position the oxygen...

Страница 14: ...326 AC POWER SUPPLY WITH TWO PINS The Eclipse 5 includes a universal AC Power Supply for use at home or wherever standard AC power is available To connect to an AC power source be sure the AC Power Su...

Страница 15: ...le from the vehicle s DC power outlet WARNING DO NOT ALLOW EITHER THE AIR INTAKE OR THE AIR OUTLET VENTS TO BECOME BLOCKED THIS CAN CAUSE THE OXYGEN CONCEN TRATOR TO OVERHEAT AND IMPAIR PERFORMANCE WA...

Страница 16: ...RES OR WITH A FULLY CHARGED OR COMPLETELY DISCHARGED BATTERY CAN DEGRADE ITS OVERALL BATTERY LIFE DO NOT ATTEMPT TO OPEN THE BATTERY THERE ARE NO SERVICEABLE PARTS INSIDE THE BATTERY KEEP BATTERIES AW...

Страница 17: ...Note Operating the Oxygen Concentrator outside of its normal operating temperature range can affect performance and decrease battery run time and or increase battery charge time Refer to the Specific...

Страница 18: ...ot detected The device will continually check for an inspiratory effort every 15 seconds Once an inspiration is detected the device reverts to measured bolus delivery To exit Pulse Dose Mode and engag...

Страница 19: ...7 feet 2 1 m of tubing PN 5408 SEQ For any additional recommended accessories please see the Accessories Catalog PN ML LOX0010 available on www CAIREmedical com A firebreak is required for use with an...

Страница 20: ...he yellow light is on Battery symbol is flashing on LCD The Eclipse 5 power cartridge voltage may be low 1 Plug Eclipse 5 into AC power supply and recharge or replace power cartridge with fully charge...

Страница 21: ...battery CHARGER ON OR OFF ON 1 alarm beep repeat every 20 seconds 001 Ambient pressure fail AMB PRESS ON OR OFF ON 1 alarm beep repeat every 20 seconds 004 O2 85 After warm up O2 LOW OFF ON 1 alarm be...

Страница 22: ...sformer as well 3 Ensure that the power verification lights are on There will be a green light on the power supply box There will be a green light on the Eclipse 5 concentrator control panel in the sh...

Страница 23: ...be a green light on the Eclipse 5 concentrator control panel in the shape of a power cord 4 If power verification LEDs are not lit remove all connections of the power supply for 20 seconds and reconne...

Страница 24: ...r in warm tap water using a mild soap detergent solution 3 Rinse the filter thoroughly with clear tap water and squeeze out the excess water 4 Allow the filter to air dry 5 Reinsert the filter in the...

Страница 25: ...BASED PROD UCTS CONCENTRATED CHLORINE BASED PROD UCTS ETHYLENE CHLORIDE AND OIL BASED PRODUCTS PINE SOL LESTOIL THESE ARE NOT TO BE USED TO CLEAN THE PLASTIC HOUS ING ON OXYGEN CONCENTRATOR AS THEY CA...

Страница 26: ...ssions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B The device is suitable for use in all establishments including domestic es...

Страница 27: ...mmended that the Eclipse 5 is powered from an uninterruptible power supply UPS Power frequency magnet ic field IEC 61000 4 8 30A m Power frequency magnetic fields should be at levels characteristic of...

Страница 28: ...28 ENG PN 21182453 C User Manual Eclipse 5 NOTES...

Страница 29: ...PN 2118245 C User Manual 29 ENG Eclipse 5 NOTES...

Страница 30: ...30 ENG PN 21182453 C User Manual Eclipse 5 NOTES...

Страница 31: ...2 3 3 4 5 5 6 7 8 Table is based on 20 Breaths Per Minute for Pulse Dose Mode only 7 1 2 2 3 3 4 5 5 6 6 8 1 2 3 3 4 5 6 6 7 8 9 1 2 3 4 5 6 7 8 8 9 Table is based on 15 Breaths Per Minute for Pulse...

Страница 32: ...2018 Chart Industries Chart Industries reserves the right to discontinue its products or change the prices materials equipment quality descriptions specifications and or processes to its products at a...

Отзывы: