700 CLD / O2
5-04-2017
87
Diagnostic Values
→
→
→
or
→
Temperatures Screen
The Diagnostic Values screens allow the operator to check analyzer temperatures,
sample and air pressures, EPC voltage percentages and flows.
These important
screens are accessed by pressing
from the Diagnostics menu.
The first screen that appears is the Temperatures screen. The Temperatures screen
displays the current temperature and low and high alarm limits for key analyzer
components.
As indicated at the bottom of the screen, use the left and right arrow keys
to
scroll to different screens. The current screen will be highlighted (Temps in the example).
The Temperatures, Pressures and EPC Voltage Percent screens include the current device
Values and the Low and High Alarm Limits. For example, if the analyzer’s diode
temperature drops lower than -5.50
°
C or exceeds -4.50
°
C, an alarm will be triggered and
displayed at the bottom of the
Measure Screen
.