CAI 700 CLD Series Скачать руководство пользователя страница 3

700 CLD / O2 

 

5-04-2017 

 
 

 

 

Range Setup

 ...................................................................................................................... 81

 

   Range Limits ........................................................................................................................... 82

 

   Auto Range On/Off ................................................................................................................. 84

 

   AutoRange Switch Points ....................................................................................................... 85

 

Diagnostics

 ........................................................................................................................ 86

 

   Diagnostic Values ................................................................................................................... 87

 

   Raw Values Display ................................................................................................................ 89

 

   I/O Status ................................................................................................................................. 91

 

      Analyzer Digital Outputs ..................................................................................................... 92

 

      Analyzer Digital Inputs ........................................................................................................ 93

 

      Programmable Digital Outputs............................................................................................. 94

 

   Status Line ............................................................................................................................... 95

 

Setup Menu

 ....................................................................................................................... 96

 

   Measure Settings ..................................................................................................................... 97

 

Averaging Time ......................................................................................................................... 98

 

NO/NO

x

/NO

2

 Mode Times ........................................................................................................ 99

 

NOx Correction Factors ........................................................................................................... 100

 

   Output Settings ...................................................................................................................... 102

 

      Programmable Analogs ...................................................................................................... 103

 

         Output Assignments ........................................................................................................ 104

 

         Output Scaling ................................................................................................................. 106

 

         Output Adjustments ........................................................................................................ 108

 

      Programmable Digitals ....................................................................................................... 110

 

         Output Assignments ........................................................................................................ 111

 

         Output Hold/Clear ........................................................................................................... 114

 

         Output Test ...................................................................................................................... 115

 

   TCP/IP Parameters ................................................................................................................ 117

 

   Data Logging Time ............................................................................................................... 118

 

   Auto Start Settings ................................................................................................................ 119

 

   Clock Settings ....................................................................................................................... 121

 

Alarms Menu

 .................................................................................................................. 122

 

   Current Alarms ...................................................................................................................... 123

 

   Alarm Log ............................................................................................................................. 124

 

   Alarm Limits ......................................................................................................................... 126

 

      Temperatures ...................................................................................................................... 127

 

      Pressures ............................................................................................................................. 128

 

      EPC Voltages ..................................................................................................................... 129

 

      Concentrations.................................................................................................................... 130

 

   Alarm Display On/Off ........................................................................................................... 131

 

Service Menu

 .................................................................................................................. 132

 

   Linear Coefficients ................................................................................................................ 133

 

   Factory Service...................................................................................................................... 135

 

Security

 ........................................................................................................................... 136

 

   Operator Levels ..................................................................................................................... 137

 

   Change Password .................................................................................................................. 138

 

   Reset Password...................................................................................................................... 139

 

Communication and Interface

 ............................................................................................. 140

 

Analog and Digital Interface

 ............................................................................................ 141

 

Serial Interface

 ................................................................................................................ 145

 

AK Protocol

 .................................................................................................................... 146

 

Содержание 700 CLD Series

Страница 1: ...700 CLD O2 5 04 2017 1 CAI 700...

Страница 2: ...x NO2 Mode 50 Automated Calibration 51 Initiate Sequenced Cal 53 Initiate Sequenced Check 55 Cal Gas Concentrations 56 Calibration Setup 57 Auto Calibration Schedule 58 Auto Calibration Parameters 60...

Страница 3: ...put Assignments 104 Output Scaling 106 Output Adjustments 108 Programmable Digitals 110 Output Assignments 111 Output Hold Clear 114 Output Test 115 TCP IP Parameters 117 Data Logging Time 118 Auto St...

Страница 4: ...nds 153 Control Commands 164 Configuration Commands 168 Modbus Protocol 174 MBAP Description 174 MBAP Header Description 175 Modbus Command Function Codes 176 Modbus Map 193 01H Single Read Coil 193 0...

Страница 5: ...The CAI 700 CLD Analyzer is an exceptionally accurate chemiluminescent CLD gas analyzer designed for measuring oxides of nitrogen in gas samples It also includes a paramagnetic oxygen channel Features...

Страница 6: ...In the NOx mode NO plus NO2 is determined as above however the sample is first routed through the internal NO2 to NO converter which converts the NO2 in the sample to NO The resultant reaction is dire...

Страница 7: ...to 25 PSIG Sample flows through the sample EPC valve the NO NOx solenoid valve the sample capillary and into the reaction chamber The sample pressure is factory set to approximately 3 85 PSIG Standar...

Страница 8: ...hour period Zero and Span Adjustment Via front panel TCP IP or RS 232 NH3 HCN and SO2 Effect Not detectable with 100 ppm CO2 Effect Less than 2 with 10 CO2 H2O Effect Less than 1 with 1 H2O Flow Cont...

Страница 9: ...ction chamber Display 3 x 5 backlit LCD Sample Temperature Up to 50 C non condensing sample Ambient Temperature 5 to 40 C Ambient Humidity Less than 90 RH non condensing Warm up Time 1 hour typical Fi...

Страница 10: ...n important detail A CAUTION safety alert appears with information that is important for protecting your equipment and its performance A WARNING safety alert appears with information that is important...

Страница 11: ...CAUTION or WARNING statement Some or all of the above symbols may appear in this manual or on the equipment This manual should be consulted whenever one of these symbols is encountered on the equipmen...

Страница 12: ...ping container your analyzer arrives in The shipping container and packaging are specially designed to protect the analyzer in transport If you ever need to return the analyzer to CAI for repair or an...

Страница 13: ...functions 4 TCP IP connection to connect network connector 5 Serial connection to connect serial connector cable 6 Zero Gas inlet for feeding hydrocarbon free zero air to the analyzer Only present wi...

Страница 14: ...ble gases 4 The instrument must not be subject to severe vibration If severe vibration is present use isolation mounts 5 The instrument is designed for rack mounting Optional rack mount slides are ava...

Страница 15: ...s for the various output selection options Shielded wiring is recommended for output signals Recommended Gases Zero calibration for the 700 CLD requires ultra high purity nitrogen UHP N2 or calibratio...

Страница 16: ...e ammonia scrubber be purchased and installed on the sample line prior to the sample entering the analyzer Gas Connections If the calibration gases are not connected to calibration inlets on the back...

Страница 17: ...resence of Corrosive Gases The useful service life of the instrument will be shortened if high concentrations of corrosive gases such as Cl2 SO2 F2 HCl etc are present in the sampled gas Gas Temperatu...

Страница 18: ...tion gas cylinder pressures should be set at 25 PSIG for delivery into the optional zero and span inlets located on the rear panel CAUTION If the analyzer contains an optional internal sample pump the...

Страница 19: ...check the power source and the fuse If the problem persists call CAI Technical Support Refer to the Using the Keypad section and review the complete Operator s Manual for detailed instructions for pro...

Страница 20: ...ng box was retained the analyzer should be stored in the box in the packing material supplied If the original box is not available and another appropriate box cannot be obtained the analyzer can be pl...

Страница 21: ...values This is why each number key includes both the larger Function number at the top for example F1 and the smaller number underneath for numeric value for example 1 The Light key is used to turn th...

Страница 22: ...splayed in the bottom right corner of the screen when the analyzer is in Numeric Entry mode An F is displayed when the analyzer is being used for Function mode In Function mode the arrow keys move the...

Страница 23: ...cess the Main Menu to quickly find any screen Menu Main Menu F1 Measure F2 Measure Mode F1 NO Mode F2 NOX Mode F3 NO NOX NO2 Mode F4 Wet Dry with option F3 Menus F1 Calibration F2 Range Setup F3 Diagn...

Страница 24: ...Pump Valves F4 Auto Cal Check F5 AutoCal Timing F6 Deviation Limits F7 Analog Hold on Cal F5 Calibration Factors F1 Manual Deviations F2 Auto Cal Deviations F3 Offset Gain Factors F6 Reset Factory Se...

Страница 25: ...up Menu Main Menu F3 Menus F2 Range Setup F1 Range Limits F2 AutoRange On Off F1 AutoRange On F2 AutoRange Off F3 AutoRange Switch Points Security Level 3 FACTORY Security Level 2 SETUP Security Level...

Страница 26: ...3 Menus F3 Diagnostics F1 Diagnostic Values F1 Temps Press EPC Flows F2 Raw Value Display F3 I O Status F1 Analyzer Digital Outputs F2 Analyzer Digital Inputs F3 Programmable Digital Outputs F4 Status...

Страница 27: ...F1 Programmable Analogs F1 Output Assignments F2 Output Scaling F3 Output Adjustments F2 Programmable Digitals F1 Output Assignments F2 Output Hold Clear F1 Hold Outputs F2 Clear Outputs F3 Output Tes...

Страница 28: ...nt Alarms F2 Alarm Log F3 Alarm Limits F1 Temperatures F2 Pressures F3 EPC Voltages F4 Concentrations F4 Alarm Display On Off F1 Turn Alarm Display On F2 Turn Alarm Display Off Security Level Legend S...

Страница 29: ...Menu Main Menu F3 Menus F6 Service F1 Linear Coefficients F1 Range 1 F2 Range 2 F3 Range 3 F4 Range 4 F2 Factory Service CAI use Only Security Level Legend Security Level 1 STANDARD Security Level 2...

Страница 30: ...n Menu F3 Menus F7 Security F1 Operator Levels F1 Standard Functions F2 Setup Functions F3 Factory Functions F2 Change Password F3 Reset Password Security Level Legend Security Level 1 STANDARD Securi...

Страница 31: ...function keys All software functions of the 700 CLD Analyzer can be reached via the menu above from the Main Menu screen Operation starts by pressing the Menu key to bring up the Main Menu Use the Ar...

Страница 32: ...Please review the descriptions below corresponding with the callouts on the illustration above to familiarize yourself with the Measure Screen Screen Name The name of the active screen the Analyzer i...

Страница 33: ...analyzer s full scale value of the range currently in use Current Time Active Alarms Scrolls between Time and Date and any active alarms Keypad Status Indicates how the keypad input is currently bein...

Страница 34: ...Manual Calibration menu Menus The Menus screen is the starting point for advanced setup and functions Standby When the analyzer is in Standby mode it closes all valves and turns off the analyzer s opt...

Страница 35: ...O NOx or NO NOx NO2 The Measure Mode menu is accessed by pressing the key on the Main Menu This menu will affect how the analyzer operates and what is displayed in the Measure screen Press to set the...

Страница 36: ...ode menu In NO mode the sample gas does not pass through the analyzer s NOx converter The final reading is NO only NOx Mode To change to the NOx mode press while in the Measure Mode menu In NOx mode t...

Страница 37: ...set on the NO NOx NO2 Mode Times screen All NO and NOx readings are displayed as averaged values The cycle begins with the NO reading When the NO reading is completed the analyzer switches to the NOx...

Страница 38: ...sed by pressing the key on the Main Menu This screen includes the Model and Serial Number of your analyzer for easy identification if you are discussing your analyzer with CAI factory settings for sam...

Страница 39: ...nual menu is accessed by pressing the key on the Main Menu The current setting Remote Mode is displayed in the upper right hand corner of the screen Example SREM The analyzer can be controlled remotel...

Страница 40: ...Standby or When the analyzer is in Standby Mode the pump is turned off and the solenoid valves are closed The CAI logo is displayed along with the Serial Number Standby mode is accessed by pressing th...

Страница 41: ...tup and Diagnostics From the Main Menu press to bring up the Menus screen Press to access the Calibration menus Press to access the Range Setup menu Press to access the Diagnostics menus Press to acce...

Страница 42: ...ation menu includes important features including basic setup for both manual and automated calibration Preparing the Analyzer for Calibration NOTE If you are changing the analyzer s factory settings C...

Страница 43: ...calibration is performed a span calibration or check should be done afterward The analyzer has two separate manual calibration menus one for NO or NOx modes and a second menu for NO NOx NO2 mode The...

Страница 44: ...ngle Mode operation the analyzer can only be calibrated for one mode The Manual Calibration menu is accessed by pressing from the Calibration menu Note For O2 calibration see examples for NOx Press to...

Страница 45: ...m In the upper right corner of the screen you will see a status line that indicates how the calibration gas is being introduced into the analyzer In this case Cal Via Pump is displayed The other optio...

Страница 46: ...ero portion of the calibration should now be complete If the calibration was successful the screen will say Saved Current above the concentration value Example of a successful calibration If the calib...

Страница 47: ...rrect gas was introduced into the analyzer 2 Verify the Diagnostic Values while flowing gas 3 Check Zero Gas Deviations under Manual Deviations 4 Check Maximum Calibration Error under Calibration Setu...

Страница 48: ...e press and change the concentration to match it Press again to close the span gas concentration field Make sure the analyzer is in the range you wish to calibrate Use the Up Down arrows to go to the...

Страница 49: ...viation Limits See the Manual Zero Calibration section for examples of screens showing successful and unsuccessful calibrations If the span calibration is unsuccessful check the following 1 Make sure...

Страница 50: ...ess the Manual Zero Calibration screen For details on zero calibration see Manual Calibration NO or NOx Mode Press to access the Manual Span Calibration screen For details on span calibration see Manu...

Страница 51: ...ation menu Sequenced means that the flow times of both zero and span gases are controlled using a timer See AutoCal Timing located in the Calibration Setup menu NOTES An automated calibration should n...

Страница 52: ...calibration can be preset to include the desired interval for recurring analyzer initiated calibrations This requires additional setup Automatic calibration of multiple ranges is also possible See Ca...

Страница 53: ...he sequenced calibration is initiated it will calibrate the analyzer in the current mode and range In this case the NOx mode is shown near the upper left corner of the screen A sequenced calibration h...

Страница 54: ...7 Purging With Sample Introduces sample gas back into the analyzer and clears out any remaining gases so the current measurements will not be affected by any residual calibration gases After these ste...

Страница 55: ...ation Menu screen pressing initiates a sequenced calibration check Rather than initiating a calibration it checks the validity of your most current calibration A sequenced calibration check performs a...

Страница 56: ...Menu The Cal Gas Concentrations screen displays the range identification the changeable span gas value and the full scale value set for that range Using the Up Down arrows move the highlighted field...

Страница 57: ...e Calibration Setup menu apply to Automated Calibration The following also apply to Manual Calibration Cal Pump Valves Auto Cal Check Deviation Limits and Analog Hold on Cal All settings should be ver...

Страница 58: ...addition to the Start Time and Date the Scheduled Calibration interval in the example scheduled every 3 days can be changed by the operator The Auto Calibrations screen is accessed by pressing from t...

Страница 59: ...g Hourly Daily or Weekly to save your changes If this is not done the selected changes will not be made and the analyzer will revert to the previous settings To change Timed Auto Calibration to on or...

Страница 60: ...field you intend to change Press to open the field and change the parameter Press again to close the field after you have made your change The first changeable parameter is the Range to be calibrated...

Страница 61: ...Zero gas only or All calibration gases zero and span gases Press to open the field and change the parameter using the up down arrows Press to close the field Press to save your settings Once your cha...

Страница 62: ...on Setup Menu Please note that the Calibration Setup menu will display the current settings on the right side of the screen after the ellipsis Example Cal Pump Valves Pump NOTE If the analyzer is equi...

Страница 63: ...to set the analyzer parameter to Calibrate This setting will be saved and the analyzer will return to the Calibration Setup menu The Calibration Setup menu shows the current setting on the right side...

Страница 64: ...gate between parameters use the up or down arrow to move the highlight to the field you intend to change Press to open the field and change the value seconds Press again to close the field after you h...

Страница 65: ...ng time is factory set at 10 seconds and cannot be changed by the operator 3 Verifying Time during this time the measured value is checked to make sure it does not deviate outside the upper or lower l...

Страница 66: ...the maximum acceptable error limits of the zero and span gases for both manual and sequenced calibration To access the Deviation Limits menu press from the Calibration Setup menu Press to set or view...

Страница 67: ...to accept a calibration To access the Maximum Calibration Error screen press from the Deviation Limits menu To navigate between fields use the up or down arrow to move the highlight to the field you i...

Страница 68: ...ed calibration To access the Maximum Verifying Error screen press from the Deviation Limits menu To set the allowable tolerances for different ranges use the up or down arrow to move the highlight to...

Страница 69: ...g Hold On Cal menu press to turn Analog Hold On which will hold the analog outputs at the last measured value You will return to the Calibration Setup menu The Calibration Setup menu shows the current...

Страница 70: ...ck and view changes from the factory and previous calibrations To access the Calibration Factors menu press from the Calibration menu Press to view the Manual Calibration Deviations menu Press to view...

Страница 71: ...nual Deviations menu allows the operator to view the Zero and Span Deviations from manual calibrations Press from the Calibration Factors menu to access the Manual Deviations menu Press to view Zero G...

Страница 72: ...Zero Gas Deviations screen Absolute Zero Gas Deviation is the zero gas content calculated by the factory polynomial related to the calibrated range limit Relative Zero Gas Deviation is the current de...

Страница 73: ...view the Span Gas Deviations screen Absolute Span Gas Deviation is span gas bottle value minus span gas value calculated by the factory polynomial related to the calibrated range limit Relative Span G...

Страница 74: ...either zero or span verifying deviations The verifying deviations are taken during the verifying stage of sequenced and auto calibrations Press from the Calibration Factors menu to access the Auto Ca...

Страница 75: ...nu to view the Verifying Zero Deviations screen Measured Value is the averaged concentration during the Verifying Zero stage of sequenced and auto calibrations Variance is the difference of the measur...

Страница 76: ...w the Verifying Span Deviations screen Measured Value is the averaged concentration during the Verifying Span stage of sequenced and auto calibrations Variance is the difference of the measured value...

Страница 77: ...in Offset or Gain will provide insight into changes in analyzer performance Press from the Calibration Factors menu to access the Offset Gain Factors screen Offset is the difference between factory ze...

Страница 78: ...f resetting the Offsets and Gains or both Factory Linear Coefficients and Offsets and Gains for all calibrated ranges Resetting factory settings will not affect any other operator changed parameters P...

Страница 79: ...default settings 0 and 1 respectively and brings you to this confirmation screen Offset and Gain factors are created when the analyzer is zeroed and spanned If the Offsets and Gains are reset the anal...

Страница 80: ...ients Offset and Gain Factors and NOx Correction Factors to factory default settings and brings you to this confirmation screen NOTE After resetting Linear Coefficients the analyzer must be zeroed and...

Страница 81: ...Menus screen press to access the Range Setup menu Press to view or change Range Limits Press to access the Auto Range On Off menu In either case you will return to the Range Setup menu The Range Setu...

Страница 82: ...down arrows to move the highlight to the field you intend to change Press to open the field to change the value in ppm Press again to close the field Press to save your changes To initiate the saved c...

Страница 83: ...t two ranges set Range 1 to the lowest value Range 2 to the highest value and the others to zero 5 If new ranges are saved the Auto Range Switch Points will be set to default percentages of range limi...

Страница 84: ...en The current Auto Range status appears in the upper right corner of the screen Press to turn Auto Range On This function allows the analyzer to automatically change ranges without the presence of an...

Страница 85: ...wn Switch Point is 90 of the previous range s Up Switch Point To change the Auto Range Switch Points use the up down arrows to move the highlight to the field you intend to change Press to open the fi...

Страница 86: ...the Diagnostic Values screen It allows you to check analyzer temperatures pressures EPC voltage percentages and flows Press to access the Raw Values Display menu Press to access the I O Status menu Y...

Страница 87: ...reen displays the current temperature and low and high alarm limits for key analyzer components As indicated at the bottom of the screen use the left and right arrow keys to scroll to different screen...

Страница 88: ...h alarm limits in PSIG EPC Voltage Percent Screen The EPC screen displays the percentage of EPC voltage being supplied to the EPC valve Flows Screen The Flows screen displays the current flow of sampl...

Страница 89: ...e the calibration curve from which the Raw Concentration and Measured Concentration are derived The 0 512 volts is equal to 0 ppm and 4 512 is equal to the four factory set range limits Example of sta...

Страница 90: ...cated above the range at the bottom right of the screen To return to the Measure mode press F4 again Meas will be indicated above the range on the screen Press to open the Span valve if your analyzer...

Страница 91: ...he analyzer s digital outputs or digital inputs open or closed To access the I O Status menu press from the Diagnostics menu Press to view the status of the analyzer s standard digital outputs Press t...

Страница 92: ...at are used to control the digital output function In the example M7 M is the Main Connector on the analyzer s back panel and 7 is the Pin Number on that connector The Status becomes closed when the f...

Страница 93: ...ent Analyzer Digital Input screens The Pin column indicates the connector and the pin number that are used to control the function In the example M12 M is the Main Connector on the analyzer s back pan...

Страница 94: ...s indicated at the bottom of the screen press the left and right arrow buttons to view the next or previous page of Programmable Digital Output statuses The column key is as follows Aux Pin Auxiliary...

Страница 95: ...tocol information See AK Protocol The current status is shown in the upper right corner of the Status Line menu Example above OFF From the Diagnostics menu Press to select On or Off Press from the Sta...

Страница 96: ...the operator to view or change averaging times NO NOx NO2 mode times and NOx correction factors Press to access the Output Settings menu The Programmable Analog and Programmable Digital outputs can b...

Страница 97: ...rs Averaging Time NO NOx NO2 Mode Times and NOx Correction Factors The Measure Settings menu is accessed by pressing from the Setup menu Press to view or change the Averaging Time of the measured conc...

Страница 98: ...menu press to access the Averaging Time screen The Averaging Time is a sliding average As shown above it can be set from 0 60 seconds Press to open the field to change the time After making your chan...

Страница 99: ...and NOx cycles Any gas remaining in the sample stream from the previous mode is purged before integration into the next mode All NO and NOx readings are displayed as averaged values and the sample rea...

Страница 100: ...normal operation To properly adjust for any offsets all four steps need to be done in order from through There is only one set of Offset and Correction Factors for all ranges Calibrating in NOx single...

Страница 101: ...on does not match press and change the concentration to match Once the concentration stabilizes press to set the span Press to return to the NOx Correction Factors menu Press to set the analyzer in NO...

Страница 102: ...Output Settings menu is accessed by pressing from the Setup menu Press to see the Programmable Analogs menu which allows the operator to view or change the analog Output Assignments Output Scaling or...

Страница 103: ...s accessed by pressing from the Output Settings menu Press to view or reassign the four programmable analog output signals Press to view or set the output scaling of programmable analog output signals...

Страница 104: ...s to access the Output Assignments screen Use the Up Down arrows to highlight the field you intend to change Press to open the field and use the Up Down arrows to change it to the desired signal Press...

Страница 105: ...will be a live reading In NO NOx NO2 mode the output will be a read and hold SamplePres Sample pressure psig AirPres Air pressure psig OvenTemp Oven temperature C Heated version ConvTemp Converter te...

Страница 106: ...n output for a specific concentration From the Programmable Analogs menu press to access the Output Scaling screen Use the Up Down arrows to highlight the field you intend to change Press to open the...

Страница 107: ...t settings allow the output voltage to follow the range limits Example If Range 1 is set to 10 ppm and Range 2 is 100 ppm in Range 1 10 ppm will 10 VDC and in Range 2 100 ppm will 10 VDC 3 When the an...

Страница 108: ...back to Measure For information on analog output connections see Analog and Digital Interface From the Programmable Analogs menu press to access the Output Adjustments screen Output Type can be set f...

Страница 109: ...e FS column and press until the FS column value reads 100 0 Then use the Left Right arrows to highlight the Gain column Press to open the field Make a small adjustment and press again to close the fie...

Страница 110: ...is accessed by pressing from the Output Settings menu From the Programmable Digitals menu Press to assign any of the 15 programmable digital outputs to a specific alarm or status Press to set the prog...

Страница 111: ...ments screen Use the left and right arrow buttons to highlight the field you intend to change Press to open the field and use the up or down arrow button to change the signal Press again to close the...

Страница 112: ...d R2NC Range 2 not calibrated R3NC Range 3 not calibrated R4NC Range 4 not calibrated O2NC O2 not calibrated Conc1 Concentration Alarm 1 Conc2 Concentration Alarm 2 O2ADC O2 ADC Alarm O2C1 O2 Concentr...

Страница 113: ...Pressure Air Pressure Oven Temperature Converter Temperature Pump Temperature Diode Temperature Cell Temperature Dryer Temperature O2 Temperature Sample EPC Voltage Air EPC Coil Voltage ADC Over Range...

Страница 114: ...Output Hold Clear menu is accessed by pressing from the Programmable Digitals menu Press to hold programmed alarm outputs until they are manually cleared Press to set the outputs to automatically cle...

Страница 115: ...ressing from the Programmable Digitals menu The Output Test Warning screen first appears asking if you wish to continue the output test If you proceed with the test you must reboot the analyzer to exi...

Страница 116: ...e the state of the output open or closed Press again to change the state back To test all the outputs at once press to change the state of all the outputs they are all Open in the example Press again...

Страница 117: ...et Protocol specifies the addressing scheme Most networks combine IP with TCP establishing a virtual connection between destination and source The IP address Netmask and Gateway may be defined by the...

Страница 118: ...n this feature on These files can only be accessed by CAI To access the Data Logging screen press from the Setup menu To turn Data Logging on press to open the field and change the Logging Interval fr...

Страница 119: ...he analyzer was last in The operator can set the following parameters Auto Startup Turn the Auto Startup function on or off Wait for min Allow for the time it takes for the analyzer to warm up before...

Страница 120: ...ting use the Up Down arrows to move the highlight to the setting you wish to change Press to open the field to change the value After making your change press again to close the field Press to save yo...

Страница 121: ...alibrations and data time stamping The Clock Settings screen is accessed by pressing from the Setup menu To change a setting use the Up Down arrows to move the highlight to the setting you wish to cha...

Страница 122: ...e alarms that are currently active Press to access the Alarm Log The operator can view a log of the last 40 alarms Press to access the Alarm Limits menu The operator can view or change the existing al...

Страница 123: ...Range AirP Air Pressure AUnd ADC Under Range OvenT Oven Temperature R1NC Range 1 not calibrated ConvT NOx Converter Temperature R2NC Range 2 not calibrated PumpT Pump Temperature R3NC Range 3 not cali...

Страница 124: ...view the analyzer s last 40 alarms and their current statuses Press to access the Alarm Log screen from the Alarms menu Press to view the next page Press to return to the previous screen Screen Name...

Страница 125: ...R2NC Range 2 not calibrated PumpT Pump Temperature R3NC Range 3 not calibrated DiodT Diode Temperature R4NC Range 4 not calibrated CellT Cell Temperature O2NC O2 not calibrated DryT Dryer Temperature...

Страница 126: ...Temperatures screen It allows the operator to set the upper and lower temperature alarm limits Press to access the Pressures screen It allows operator to set the upper and lower alarm limits for sampl...

Страница 127: ...Limits menu press to access the Temperatures screen Use the left or right arrow to highlight the alarm limit you wish to change Press to open the field to change the value After making a change press...

Страница 128: ...Limits menu press to access the Pressures screen Use the left or right arrow to highlight the alarm limit you intend to change Press to open the field to change the value After making a change press a...

Страница 129: ...menu press to access the EPC Voltage Alarm Limits screen Use the left or right arrow to highlight the alarm limit you intend to change Press to open the field to change the value After making a chang...

Страница 130: ...he concentration alarm limit you intend to change Press to open the field to change the value After making a change press again to close the field After the concentration limit is set specify whether...

Страница 131: ...ting only the display will be turned off The scrolling alarms will be replaced with the CAI phone number when an alarm is active If there are no alarms only the date and time will be displayed Press f...

Страница 132: ...and CAI Service The Service menu provides access to operator level linear coefficients and CAI factory service From the Menus screen press to access the Service menu Press to view or change operator...

Страница 133: ...h order curve From the Service menu press to access the Linear Coefficients screen NOTES Changing linear coefficients can compromise the analyzer s accuracy and ability to function properly To reset F...

Страница 134: ...highlight the coefficient you intend to change Press to open the field use the left and right arrow buttons to position the cursor and use the Up Down arrows or number keys to make your change Press a...

Страница 135: ...ce The Factory Service menus can only be accessed by California Analytical Instruments The best way to contact CAI with a support question is to fill out our brief Technical Support Form at http www g...

Страница 136: ...s screen press to access the Security menu Press to set the operator s access level The Operator Levels menu allows the operator to choose either Standard Functions or advanced Setup Functions Press t...

Страница 137: ...and functions Press to set the operator level to Standard Functions for basic operation and calibration No password is required Press to set the operator level to Setup Functions for advanced operato...

Страница 138: ...hange the Setup Functions password from the factory preset 222 to a new password From the Security menu screen press to access the Operator Levels screen To change the password press to open the field...

Страница 139: ...the Reset Password screen To return the password to the original factory password you must enter WIZARD in the field Press to open the field and bring up a cursor Use the Up arrow to scroll to W Use t...

Страница 140: ...or specific voltage range and signal assignments are assignable from the Setup menu See the following tables for pin numbers and functions Serial Interface The 9 pin Serial Interface connector provid...

Страница 141: ...be used as the return line for the four analog outputs See Programmable Analog Output Adjustments Digital Output There are two types of digital outputs Static Digital Outputs Permanently assigned to...

Страница 142: ...l Ground Main Pin 6 Note that some lines require only momentary operation and some selection lines are intended to be held continuously DRV The calibration gas valve drivers are application specific a...

Страница 143: ...Digital Output NO Relay to Digital Ground 12 Set Auto Range Digital Input Hold to Ground in Remote 13 Control Range 1 Digital Input Hold to Ground in Remote 14 Control Range 2 Digital Input Hold to G...

Страница 144: ...ay 5 Digital Output Uses Relay RTN 2 14 User Defined NO Relay 6 Digital Output Uses Relay RTN 2 15 User Defined NO Relay 7 Digital Output Uses Relay RTN 2 16 User Defined NO Relay 8 Digital Output Use...

Страница 145: ...used to connect a master computer with the following pin assignment Figure 0 1 Serial Interface Pin 3 Txd transmit Pin 2 Rxd receive Pin 5 Gnd ground Interface Specifications Baud Rate 9600 4800 2400...

Страница 146: ...mber is usually 0 For delimiting the command parameters from the channel number another blank follows This may be followed by command parameters with variable lengths Every command ends with the ETX E...

Страница 147: ...e analyzer Sending the analyzer characters not framed as AK commands should be prevented as the menu could recognize these characters and cause unplanned changes in configuration The AK command was ty...

Страница 148: ...g ASTF 7th Byte Blank 8th Byte K 9th Byte 0 10th Byte Blank D AK instruction parameters lengths variable A AK instruction parameters lengths variable T AK instruction parameters lengths variable A AK...

Страница 149: ...h Byte Blank 8th Byte K 9th Byte 0 10th Byte Blank D AK acknowledgement parameters lengths variable A AK acknowledgement parameters lengths variable T AK acknowledgement parameters lengths variable A...

Страница 150: ...t Explanation f Analyzer does not recognize the instruction sent xxxx f BS Analyzer is busy with another function xxxx f SE Syntax error within command parameters or incomplete command xxxx f NA Reque...

Страница 151: ...y with a running function for example SLIN every arriving control command is ignored except SRES and STBY and the response message is e g SMAN 0 BS If in the SINT mode an additional SINT KO command is...

Страница 152: ...Numeric integer value a0 a4 Polynomial coefficients s Status Yyymmdd Date of format year month and day with 2 characters each and no spaces Hhmmss Time of format hour minute and second with 2 charact...

Страница 153: ...ative dev dr AANG Verifying Zero Point Deviation During Auto Calibration Command Response Description _AANG_K0 _AANG_s_M1_z z_da_dr _M2_z z_da_dr _M3_z z_da_dr _M4_z z_da_dr _02_z z_da_dr Verifying de...

Страница 154: ..._ _a16 min_a16max Sets all alarm limits numbers and descriptions 1 Sample pressure 2 Air pressure 3 Oven temperature 4 Converter temperature 5 Pump temperature 6 Diode temperature 7 Cell temperature 8...

Страница 155: ...ng of x ADUF Flows Command Response Description _ADUF_K0 _ADUF_s_z z_y y 1 Sample flow 2 Air flow _ADUF_K0_x _ADUF_s_xflow Flow of x AEFF NOx Correction Factors Command Response Description _AEFF_K0 _...

Страница 156: ...ficients Command Response Description _AFGR_K0_Mn _AFGR_s_a0_a1_a2_a3_a4 Factory coefficients for range n _AFGR_K1_M1 _AFGR_s_a0_a1_a2_a3_a4 K1 O2 option AGRD Polynomial Coefficients Command Response...

Страница 157: ...w _M2_z z_y y_x x_w w _M3_z z_y y_x x_w w _M4_z z_y y_x x_w w _02_z z_y y_x x_w w Percent Deviation of ranges M1 to M4 and O2 with O2 option z z Zero gas relative to last calibration y y Zero gas abso...

Страница 158: ...e value of channel 1 Channel 1 is O2 with option _AKON_K4 _AKON_K5 _AKON_s_z z_y y_x x_w w_t Current measure value for all 4 ranges based on their calibration offsets and gains t Timestamp 1 10 sec AM...

Страница 159: ...ARAW_s_z z_t Raw detector volts K1 for O2 with option t Timestamp 1 10 sec ARMU Raw Engineering Value Command Response Description _ARMU_K0 _ARMU_s_z z_t Raw engineering value before linearization off...

Страница 160: ...emperature 7 Check cell temperature 8 Check gas dryer temperature 9 Check O2 temperature 10 Check sample EPC 11 Check air EPC 12 Range overflow 13 ADC range overflow 14 ADC range underflow 15 Range 1...

Страница 161: ...ing gas Zero gas NO Zero gas O2 option Span gas NO Span gas O2 option Zero gas during auto cal End gas during auto cal Purging overflow State 3 SENO SNOX S2NO SNO2 NO mode NOx mode NO NOx NO2 mode NO...

Страница 162: ...ATEM_K0 _ATEM_s_z z_y y_x x 1 Oven temperature 2 Converter temperature 3 Pump temperature 4 Diode temperature 5 Cell temperature 6 Chiller temperature 7 O2 detector temperature 8 Case temperature _ATE...

Страница 163: ...Software Version Command Response Description _AVER_K0 _AVER_s_NMAIN_Z_mm dd yyyy_ NUSER_Y_mm dd yyyy_OSMSR_ X_dd mm yyyy Z NMAIN version build number Y NUSER version build number X OSMSR version bui...

Страница 164: ...able ranges _ SATK_K0_Mn _SATK_s Start automatic calibration of range n SEGA Open Valve for Span Gas Calibration Command Response Description _SEGA_K0 _SNGA_s Sets NO to span calibration mode _ SEGA_K...

Страница 165: ...gs to factory defaults Command Response Description _SFGR_K0 _SFGR_s Resets all calibration settings to their factory settings SMAN Manual Control to Control Device Manually Command Response Descripti...

Страница 166: ...e Command Response Description _SNOX_K0 _SNOX_s Sets the measurement mode to NOx only SNO2 Sets NO NOX NO2 Mode Command Response Description _SNO2_K0 _SNO2_s Sets the measurement mode to NO NOx NO2 Au...

Страница 167: ...UDP Data Streaming Command Response Description _SUDP_K0_ON _SUDP_s Starts data streaming via the UDP channel Configure the channel before starting with EUDP command _ SUDP_K0_OFF _SUDP_s Stops strea...

Страница 168: ...min_ a1 max_ _a16max _EDAL_s Sets all alarm limits numbers and descriptions 1 Sample pressure 2 Air pressure 3 Oven temperature 4 Converter temperature 5 Pump temperature 6 Diode temperature 7 Cell t...

Страница 169: ...erify time x Purge after z y x w in seconds _ EFDA_K0_ SSPL_z _EFDA_s Sets analyzer purge time to z seconds EGRD Set the Range Polynomial Coefficients Command Response Description _EGRD_K0_Mn_ A0_a1_a...

Страница 170: ...onse Description _EMBE_K0_M1_ w w_M2_x x_M3_ y y_M4_z z _EMBE_s Sets the range full scale limits EMBU Set the Upper and Lower Range Switchover for Auto Range Command Response Description _EMBU_K0_M1_...

Страница 171: ...hour minutes seconds each 2 characters no spaces ET90 Set Lowpass Filter Time Command Response Description _ET90_K0_t _ET90_s Sets lowpass filter time t filter time ETCP Set TCP IP Parameters Command...

Страница 172: ...ening the UDP connection when it should not be using the IP of the connected TCP IP client optional Data Format DATA is any number of AK commands delimited by a semicolon Replace spaces in the AK comm...

Страница 173: ...er code followed by the answer See corresponding AK command description Example Sending EUDP K0 7501 2 A AKON_KO ADUF_K0 will give the following streaming result 123 AKON 4 07 901 33 22 50 3481639460...

Страница 174: ...es such as bridges routers and gateways that use a single IP address to support multiple independent Modbus end units 2 All Modbus requests and responses are designed to allow the recipient to verify...

Страница 175: ...he request Protocol Identifier Used for intra system multiplexing The Modbus protocol is identified by the value 0 Length The Length field is a byte count of the following fields including the Unit Id...

Страница 176: ...d 0 OFF The LSB of the first data byte contains the output addressed in the query The other bits follow toward the high order end of this byte and from low order to high order in subsequent bytes If t...

Страница 177: ...puts 27 20 is shown as the byte value CD hex or binary 1100 1101 Output 27 is the MSB of this byte and output 20 is the LSB By convention bits within a byte are shown with the MSB to the left and the...

Страница 178: ...750 Manual 5 04 2017 178 Command 01 Read Coil 200 Count 1 Result 01 Command 01 Read Coil 200 Count 16 Result 55 55 hex...

Страница 179: ...starting at zero Therefore registers numbered 1 16 are addressed as 0 15 The register data in the response message are packed as four bytes per register with the binary contents right justified withi...

Страница 180: ...9A The contents of Register 0 are shown as the 4 byte values of 44 9A 52 2C hex or 1234 56789 decimal Here is an example of a request to read Register 40201 Request Response Field Name Hex Field Name...

Страница 181: ...alue Hi 40202 33 Number of Registers Hi 00 Register Value Lo 40201 33 Number of Registers Lo 06 Register Value Hi 40201 Register Value Lo 40201 Register Value Hi 40204 Register Value Lo 40204 Register...

Страница 182: ...750 Manual 5 04 2017 182 Command 03 Read four Floating Point values starting at Address 1 Result 1234 56789 0 0 1234 568 10000...

Страница 183: ...e addressed as 0 15 The register data in the response message are packed as two bytes per register with the binary contents right justified within each byte For each register the first byte contains t...

Страница 184: ...Function 04 Starting Address Hi 00 Byte Count 02 Starting Address Lo 08 Input Register 9 Hi 00 Quantity of Input Registers Hi 00 Input Register 9 Lo 0A Quantity of Input Registers Lo 01 The contents...

Страница 185: ...are addressed starting at zero Therefore the bit numbered 1 is addressed as 0 The requested ON OFF state is specified by a constant in the Bit Value field A value of 0xFF00 requests the bit to be ON A...

Страница 186: ...est Response Field Name Hex Field Name Hex Function 05 Function 05 Output Address Hi 00 Output Address Hi 00 Output Address Lo AC Output Address Lo AC Output Value Hi FF Output Value Hi FF Output Valu...

Страница 187: ...egister numbered 1 is addressed as 0 The normal response is an echo of the request returned after the register contents have been written Request Function Code 1 Byte 0x06 Output Address 2 Bytes 0x000...

Страница 188: ...ex Request Response Field Name Hex Field Name Hex Function 06 Function 06 Register Address Hi 00 Register Address Hi 00 Register Address Lo 01 Register Address Lo 01 Register Value Hi 00 Register Valu...

Страница 189: ...arting address and quantity of registers written The analyzer ignores the numbers in the register and byte count and expects four data bytes Request Function Code 1 Byte 0x10 Starting Address 2 Bytes...

Страница 190: ...ion 10 Function 10 Starting Address Hi 00 Starting Address Hi 00 Starting Address Lo 01 Starting Address Lo 01 Quantity of Registers Hi 00 Quantity of Registers Hi 00 Quantity of Registers Lo 02 Quant...

Страница 191: ...750 Manual 5 04 2017 191 Command 16 Write one floating point register at Address 40200 showing error response Not a valid address...

Страница 192: ...o read an ASCII string Request Function Code 1 Byte 0x1A Starting Address 2 Bytes 0x0000 to 0xFFFF Quantity of Registers 2 Bytes 1 Response Function Code 1 Byte 0x1A Length of String 1 Byte 0x00 to 0x...

Страница 193: ...temperature 6 Diode temperature 7 Cell temperature 8 Gas dryer temperature 9 O2 temperature with option 10 Sample EPC voltage 11 Air EPC voltage 12 Range overflow 13 ADC overflow 14 ADC underflow 15...

Страница 194: ...by 1 Measure 103 1 Zero 104 1 Span 105 1 AutoCal 106 1 Purge 107 O2 0 Standby 1 Measure 108 O2 1 Zero 109 O2 1 Span 110 O2 1 AutoCal 115 0 Via Pump 1 Via Valves 118 0 Auto Off 1 Auto On 145 NO mode 14...

Страница 195: ...118 0 Auto Off 1 Auto On 121 1 Sets current range offset to 0 0 122 1 Sets current range gain to 1 0 123 1 Sets O2 offset to 0 0 124 1 Sets O2 gain to 1 0 127 1 Sets offset of range if zero gas 128 1...

Страница 196: ...g mode 40011 NO2 concentration Switching mode 40013 NOx concentration Switching mode 40017 O2 concentration with option 40019 O2 concentration before linearization and zero span corrections 40021 O2 r...

Страница 197: ...scale 25 40133 Range 1 Auto Up 40135 Range 2 Auto Down 40137 Range 2 Auto Up 40139 Range 3 Auto Down 40141 Range 3 Auto Up 40143 Range 4 Auto Down 40201 Range 1 span gas concentration 40203 Range 2 sp...

Страница 198: ...ximum 40251 Cell temperature alarm minimum 40253 Cell temperature alarm maximum 40255 Gas dryer temperature alarm minimum 40257 Gas dryer temperature alarm maximum 40259 Sample EPC voltage alarm minim...

Страница 199: ...ressure alarm minimum 40233 Air pressure alarm maximum 40235 Oven temperature alarm minimum 40237 Oven temperature alarm maximum 40239 Converter temperature alarm minimum 40241 Converter temperature a...

Страница 200: ...04 2017 200 Register Number Contents IEEE 40267 O2 detector temperature alarm minimum 40269 O2 detector temperature alarm maximum 40287 Sample concentration alarm minimum 40289 Sample concentration a...

Страница 201: ...anted only to give reasonable service for a reasonable time which constitutes reasonable time and reasonable services shall be determined solely by CAI A complete list of such components and accessori...

Страница 202: ...RCHANTABILITY AND THAT CAI SHALL HAVE NO LIABILITY FOR SPECIAL OR CONSEQUENTIAL DAMAGES OF ANY KIND OR FROM ANY CAUSE WHATSOEVER ARISING OUT OF THE MANUFACTURE USE SALE HANDLING REPAIR MAINTENANCE OR...

Отзывы: