700 CLD / O2
5-04-2017
67
Maximum Calibration Error
→
→
→
→
→
Maximum Calibration Error is used by the operator to define the maximum
acceptable tolerances for Absolute and Relative deviations.
Each range has its own set
of Absolute and Relative tolerances. The deviations must be inside these tolerances for
the analyzer to accept a calibration. To access the Maximum Calibration Error screen.
press
from the Deviation Limits menu.
To navigate between fields, use the up or down arrow
to
move the highlight to
the field you intend to change. Press
to open the field to change the allowable
tolerance in %. Press
again to close the field. Press
to save your changes.
Absolute Deviation
is used to compare the factory-set calibration to the current
calibration.
Relative Deviation
compares the current calibration to the previous calibration.