700 CLD / O2
5-04-2017
100
NOx Correction Factors
→
→
→
→
or
→
NO
x
Correction Factors allows the operator to adjust small offsets between the NO
and NOx modes while operating in the NO/NO
x
/NO
2
mode.
Press
to access the
NO
x
Correction Factors menu from the Measure Settings menu.
Notes:
NO
x
Correction Factors is for advanced operators only and not necessary for
normal operation.
To properly adjust for any offsets, all four steps need to be done in order from
through
.
There is only one set of Offset and Correction Factors for all ranges.
Calibrating in NO
x
single mode will automatically reset the NO
x
Correction
Factor to the default settings of “0” and “1”.
Press
to set the analyzer in NO mode and enter the Manual Zero Calibration screen.
Introduce zero gas into the rear of the analyzer. Once the concentration stabilizes, press
to set the Zero. Press
to return to the NO
x
Correction Factors menu.
Current Settings