700 CLD / O2
5-04-2017
51
Automated Calibration
→
→
→
An automated calibration is a timed zero calibration immediately followed by a
timed span calibration.
The Automated Calibration menu offers two choices:
Sequenced Calibration and Sequenced Check of the existing calibration.
The Automated Calibration menu is accessed by pressing
from the Calibration
menu. Sequenced means that the flow times of both zero and span gases are controlled
using a timer. See
AutoCal Timing
located in the Calibration Setup menu.
NOTES:
An automated calibration should not be attempted before manual zero and span
calibrations have been successfully performed.
If a manually initiated sequenced calibration or sequenced calibration check is
selected, it will apply only to the range that is currently in use. (Each additional
range must be calibrated separately). This also applies if the analyzer is in auto
range.