Electrical
All wiring is connected at the rear of the analyzer. The AC power cord is connected to the
power entry as shown below:
AC Power Switch, Connector and Fuse.
NOTE:
A defective ground may affect the analyzer’s operation. Shielded wiring is
recommended for output signals.
Output Connections
See the
Analog and Digital Interface
section of this manual for instructions for the
various output selection options.
Shielded wiring is recommended for output signals.
Recommended Gases
Zero calibration for the 700 CLD requires ultra high-purity nitrogen (UHP N
2
) or
calibration-grade air, plus a span gas. The recommended span gas for this analyzer is NO
in a background of N
2
. NO in a background of air is not recommended as some of it will
convert to NO
2
.
Calibration gases can be introduced through either the calibration ports (if optional
solenoid valves have been installed) on the back of the analyzer or through the sample
Power ON/OFF
Switch
Power Entry
Fuse Panel