Section 12
Model 601/601P/602/602P/603 NDIR Analyzer
California Analytical Model 600_NDIR_Series_v10.0 CE
Operators Manual
Page 84 of 86
12.6.8.
ESYZ:
Set System Time
Command Response
Description
_ESYZ_K0_yymmdd_hhmmss
_ESYA_s
Set system time:
yymmdd: year, month, day (each 2
characters wide, no spaces)
hhmmss: hour, minutes, seconds
(each 2 characters, no spaces)
12.6.9.
ET90:
Set Lowpass Filter Time
Command Response
Description
_ET90_K0_t
_ET90_s
Set lowpass filter time:
t= filter time in seconds
12.6.10.
EDAL:
Diagnostic alarm limits
Command Response
Description
_EDAL_K0_a1.min_a1.mas_..._a12max _EDAL_s
Set all alarm limits
_EDAL_K0_x_x.min_xmax
_EDAL_s
Set alarm limits of x
Alarm Limits:
1
Flow of channel 1
7
Temperature
2
Flow of channel 2
8
Sample concentration channel 1
3
Flow of channel 3
9
Sample concentration channel 2
4
External analog in 1
10
Sample concentration channel 3
5
External analog in 2
11-13
Temperature alarm limits channel
1…3
6
Pressure
14-16
EPC voltage alarm limits channel
1…3
12.6.11.
ETCP:
Set TCP/IP Parameters
Command Response
Description
_ETCP_K0_zzz.zzz.zzz.zzz
_yyy.yyy.yyy.yyy
_xxxx
_EDAL_s
zzz= TCP/IP address
yyy= TCP/IP subnet mask
xxxx= TCP/IP port
All changes take effect after next
power on cycle
12.6.12.
EH2O
Set H
2
O correction parameters
Command Response
Description
_EH2O_Km_z.z_y.y_x.x _EH2O_s
z.z:
dry
y.y:
1st order coefficient
x.x:
2nd order coefficient
12.6.13.
ECO2
Set CO
2
correction parameters
Command Response
Description
_EH2O_Km_z.z_y.y_x.x _EH2O_s
z.z:
dry
y.y:
1st order coefficient
x.x:
2nd order coefficient