Section 1
Model 601/601P/602/602P/603 NDIR Analyzer
California Analytical Model 600_NDIR_Series_v10.0 CE
Operators Manual
Page 4 of 86
600 NDIR Quick Start Guide
1)
Plug in the analyzer and turn the power on.
2)
Connect the appropriate gas lines and vents to the analyzer.
3)
Allow the analyzer to stabilize for at least one (1) hour.
4)
During the analyzer’s stabilization period, setup the analyzer to the desired
configuration.
a) Set the analyzer’s output as desired.
i) From the Main Menu press F5 (Setup) then F7 (System Settings.)
ii) Press F3 (output assignments) to setup the output channels.
(1) Set to real time.
(2) AUX 1
(3) AUX 2
(4) AUX 3
(5) Press back to return to the system settings menu.
iii) Press F4 (output range) to setup the output ranges
(1) Set the Min and Max to zero (0). This will cause the outputs to
default to the current ranges.
(2) Set the outputs for milliamps (mA) or voltage (V) as desired.
(a) The mA setting will give 4-20 mA or 2-10V (if 500 ohm
resisters are installed.
(b) The V setting will give 0-20mA or 0-10V (if 500 ohm resistors
are installed.
Disconnect power before proceeding
5)
Connect all appropriate analog outputs.
a) Pin 1 is the output common (ground).
b) Pins 2, 3, 4, and 5 are output channels 1, 2, 3, and 4 as setup in step 4.a.ii.
c) If the output is set for voltages but there is no voltage output, you will need to
install a 500 ohm resistor between pin 1 and the channel you are trying to
measure. You will have to measure the voltage drop across this resistor.
6)
Reconnect the power and turn on the analyzer. Press F1 from the analyzer’s
main Menu to get to the measurement screen.
7)
Supply sample gas to the analyzer.
8)
The measurement screen should indicate the concentration of the sample gas. If
the concentration is incorrect, the analyzer will need to be calibrated using zero
and span calibration gases. Please reference the appropriate section of the
manual for a description on how to zero and span the analyzer.