Section 7
Model 601/601P/602/602P/603 NDIR Analyzer
California Analytical Model 600_NDIR_Series_v10.0 CE
Operators Manual
Page 42 of 86
7.5.7.
F1 Range 1-4 (Change Upper Range Limits)
In this menu, the upper range limits can be changed. The new settings are saved by
pressing MAIN or BACK. The auto range limits are automatically adapted. This means
that if the upper range limit of range 1 for example has reached 90% of the upper range
limit in the auto range mode, it is switched automatically to the second range.
Figure 7-21 Change Upper Range Limits
7.5.8.
F2 Change Auto Range Limits
Although the auto range limits are adapted automatically, it is possible to define them
manually. Up means the value when the next higher range is selected in auto range mode,
down the value when the next lower range is selected.
Figure 7-22 Change Auto Range Limits