Section 12
Model 601/601P/602/602P/603 NDIR Analyzer
California Analytical Model 600_NDIR_Series_v10.0 CE
Operators Manual
Page 76 of 86
12.4. Scan
Commands
12.4.1.
AKON:
Measured concentration value
Command Response
Description
_AKON_K0
_AKON_s_z.z_y.y_x.x_w.w
Measured concentration value of all
channels is responded
t = Timestamp (1/10 sec)
_AKON_Km
_AKON_s_z.z_t
Measured concentration value of
channel m is responded
t = Timestamp (1/10 sec)
12.4.2.
AEMB:
Get measuring range
Command Response
Description
_AEMB_K0
_AEMB_s_Mn_Mn_Mn
Current measuring range of all
channels is responded
_AEMB_Km
_AKON_s_Mn
Current measuring range of channel m
is responded
12.4.3.
AMBE:
Measuring range limit
Command Response
Description
_AMBE_K0 _AMBE_s_M1_w.w
_M2_x.x
_M3_y.y
_M4_z.z
All existing measuring range limits of
channel m are responded
_AMBE_K0_Mn
_AMBE_s_Mn_z.z
Range limit of Range Mn is responded
12.4.4.
AKAK:
Calibration gas concentrations
Command Response
Description
_AKAK_Km _AKAK_s_M1_w.w
_M2_x.x
_M3_y.y
_M4_z.z
All existing calibration gas values are
responded for selected channel m
_AKAK_Km_Mn AKAK_s_Mn_z.z
For selected channel m calibration gas
value of Range Mn is responded
12.4.5.
AMBU:
Upper and lower range switchover values for auto range
Command Response
Description
_AMBU_Km _AMBU_s_M1_w.w_W.W
_M2_x.x_X.X
_M3_y.y_Y.Y
_M4_z.z_Z.Z
Lower and upper range switchover
value of auto range are responded for
channel m
_AMBU_Km_M
n
_AMBU_s_Mn_w.w_W.W
Lower and upper range switchover
value of auto range are responded for
channel m range n