Section 10
Model 601/601P/602/602P/603 NDIR Analyzer
California Analytical Model 600_NDIR_Series_v10.0 CE
Operators Manual
Page 60 of 86
10. MAINTENANCE
Warning
All replacement parts must be as supplied and/or specified by California Analytical
Instruments. Failure to used specified parts may reduce the safety features of the
instrument or create a hazardous condition.
10.1.
Zero and Span Calibration
The zero and span levels should be checked and/or calibrated daily (or as often as required.)
10.2.
Cleaning of the Optical Bench Measuring Cell (Infrared Analyzers Only)
Dust or water droplets entering the measuring cell may cause drift due to contamination.
When it is impossible to adjust the zero level with the zero control mounted on the front panel,
check the measuring cell for contamination. If contamination is present, check the sampling
system, especially the filters, to eliminate the source of contamination. Periodic maintenance
is generally not required. Cleaning is accomplished by use of a cleaning agent (such as
isopropyl alcohol or household glass cleaner) and a non-abrasive, lint-free cloth or tissue.
10.3.
Optical Bench Configuration
Infrared analyzers may be configured with three types of optical benches: with a pipe cell; with
a block cell; or with both a pipe cell and a block cell (see Table 10-1))
Table 10-1 Optical Bench Configuration
Optical Bench Type
Illustration
Paragraph
Pipe Cell
(Cell length: 64 mm, 125 mm,
or 250 mm).
Figure 10-1
This figure illustrates the pipe
cell
10.3.1 describes the
disassembly cleaning and
reassembly
Block Cell
(Cell length: 4 mm, 8 mm, 16
mm or 32 mm)
Figure 10-2
This figure illustrates the block
cell.
10.3.2 describes the
disassembly cleaning and
reassembly
Combination
Figure 10-3
This illustrates the combination
assembly.
10.3.3 describes the
disassembly cleaning and
reassembly