4.3.20.
CAENVME_GetPulserConf
Parameters:
[in] Handle
: The handle that identifies the device.
[in] PulSel
: The pulser to configure (see CVPulserSelect enum).
[out] Period
: The period of the pulse in time units.
[out] Width
: The width of the pulse in time units.
[out] Unit
: The time unit for the pulser configuration (see
CVTimeUnits enum).
[out] PulseNo : The number of pulses to generate (0 = infite).
[out] Start
: The source signal to start the pulse burst (see
CVIOSources enum).
[out] Reset
: The source signal to stop the pulse burst (see
CVIOSources enum).
Returns:
An error code about the execution of the function.
Description:
The function permits to read the configuration of the pulsers.
CAENVME_API
CAENVME_GetPulserConf(long Handle, CVPulserSelect PulSel, unsigned char *Period,
unsigned char *Width, CVTimeUnits *Unit, unsigned char *PulseNo, CVIOSources *Start,
CVIOSources *Reset).
4.3.21.
CAENVME_GetScalerConf
Parameters:
[in] Handle
: The handle that identifies the device.
[out] Limit
: The counter limit for the scaler.
[out] AutoReset : The auto reset configuration.
[out] Hit
: The source signal for the signal to count (see
CVIOSources enum).
[out] Gate
: The source signal for the gate (see CVIOSources enum).
[out] Reset
: The source signal to stop the counter (see
CVIOSources enum).
Returns:
An error code about the execution of the function.
Description:
The function permits to read the configuration of the scaler.
CAENVME_API
CAENVME_GetScalerConf(long Handle, short *Limit, short *AutoReset,